Affiliation:
1. Stanford University Hospital, Division of Pulmonary and Critical Care Medicine, Department of Pulmonary Physiology, CA 94305
Abstract
Abstract
We evaluated a new portable instrument, the PPG StatPal II pH and Blood Gas Analysis System, designed for "point-of-care" measurements of blood gases and pH. Inaccuracy (% of target value) and imprecision (CV%) were assessed by blood tonometry and comparison with a Corning 178. Within-day results for PCO2 inaccuracy and imprecision ranged from 98.2% to 102.9% and 3.3% to 3.9%, respectively; for PO2, these were 95.5% to 102.3% and 2.3% to 3.0%, respectively. Between-day results for PCO2 inaccuracy and imprecision ranged from 99.2% to 99.3% and from 2.9% to 3.2%, respectively; for PO2, the ranges were 96.2% to 98.2% and 2.6% to 3.0%, respectively. Two PCO2 outliers (in 645 samples = 0.3%) were observed. In general, tonometry recovery, measurement stability, and pH bias results for the StatPal II and Corning 178 were comparable. We conclude that the StatPal II performs within acceptable ranges of inaccuracy and imprecision.
Publisher
Oxford University Press (OUP)
Subject
Biochemistry, medical,Clinical Biochemistry
Cited by
8 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献