Abstract
Abstract
We compared cyclosporine concentrations in whole blood as measured by HPLC and by RIA with a monoclonal antibody specific for cyclosporine with 3H- or 125I-labeled cyclosporine ligand. The 3H-RIA kit slightly underestimated cyclosporine concentrations (greater than 600 micrograms/L) in comparison with HPLC. Over a wide range of concentrations, cyclosporine measured with the 125I-RIA kit correlated well with HPLC (slope = 0.99, n = 301, r = 0.98), observed for samples from recipients of kidney, heart, or liver allografts (respective slopes: 1.01, 0.93, and 1.00). The 125I-RIA standard curve was linear to 1000 micrograms of cyclosporine per liter. Inter- and intra-assay CVs for 125I-RIA measurements of cyclosporine were less than or equal to 7%. Evidently, the 125I-RIA kit involving a monoclonal antibody specific for cyclosporine is equivalent to the HPLC assay and can replace it for therapeutic drug monitoring of cyclosporine therapy.
Publisher
Oxford University Press (OUP)
Subject
Biochemistry, medical,Clinical Biochemistry