Affiliation:
1. Department of Laboratory Medicine, Osaka City University Medical School, Japan
Abstract
Abstract
We have developed an assay system for measuring lactate in whole blood, consisting of a single-use strip of an enzyme-coated electrode and a small meter. The electrode strip is made of three plastic films: a cover sheet, a spacer, and an insulation layer printed with electrodes that are coated with lactate oxidase (EC 1.1.3.x) and ferricyanide as an electron mediator. The meter measures the magnitude of the anodic current of the reduced mediator by the enzymatic reaction and displays the lactate concentration 60 s after a blood sample (5 microL) is applied. The calibration curve was linear up to 20 mmol/L, and the between-run CVs at three concentrations were 1.7-8.4%. Lactate concentrations determined by this method (y) in blood samples from healthy individuals before and after exercise agreed with the results obtained by the conventional enzymatic method (x): y = 0.97x - 0.3, Sy/x = 0.7. This assay provides a rapid and convenient test for measuring blood lactate concentrations.
Publisher
Oxford University Press (OUP)
Subject
Biochemistry, medical,Clinical Biochemistry
Cited by
68 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献