Affiliation:
1. Organic Analytical Research Division, National Institute of Standards and Technology, Gaithersburg, MD 20899
Abstract
Abstract
Here we compare the Reference Method (I) and the Definitive Method (II) for measurement of cholesterol in serum. For pure cholesterol solutions, values by I agree with values by II, but for fresh, frozen, or lyophilized sera, values by I average 1.6% higher than values by II. We conclude that an undetected interference is associated with I. However, the observed difference does not diminish the usefulness of I as an accuracy base for cholesterol measurements, because it is clinically insignificant at present.