Abstract
Abstract
In this paper, we present initial quality-control evaluations for a prototype clinical laboratory, recommendations for clinical and technical data record-keeping, and laboratory protocols designed to minimize operator-dependent analytical error in a multi-user environment. These evaluations and recommendations are based on qualitative and semiquantitative two-dimensional electrophoresis studies of the College of American Pathologists' Reference Preparation for Serum Proteins (RPSP). We include the schematic map of RPSP and manually derived estimates of analytical precision for within-run and run-to-run performance with the standard 18 X 18 cm ISO-DALT system.
Publisher
Oxford University Press (OUP)
Subject
Biochemistry, medical,Clinical Biochemistry
Cited by
19 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献