Research and clinical translation of trilayer stent-graft of expanded polytetrafluoroethylene for interventional treatment of aortic dissection

Author:

Wang Gang12,Gao Caiyun1,Xiao Benhao2,Zhang Jie2,Jiang Xunyuan1,Wang Qunsong1,Guo Jingzhen1,Zhang Deyuan2,Liu Jianxiong2,Xie Yuehui2,Shu Chang34,Ding Jiandong1

Affiliation:

1. State Key Laboratory of Molecular Engineering of Polymers, Department of Macromolecular Science, Fudan University , Shanghai 200438, China

2. R&D Center, Lifetech Scientific (Shenzhen) Co., Ltd. , Shenzhen 518057, China

3. Department of Vascular Surgery, the Second Xiangya Hospital of Central South University , Changsha 410011, China

4. State Key Laboratory of Cardiovascular Diseases, Center of Vascular Surgery, Fuwai Hospital, National Center for Cardiovascular Diseases, Chinese Academy of Medical Science and Peking Union Medical College , Beijing 100037, China

Abstract

Abstract The aortic dissection (AD) is a life-threatening disease. The transcatheter endovascular aortic repair (EVAR) affords a minimally invasive technique to save the lives of these critical patients, and an appropriate stent-graft gets to be the key medical device during an EVAR procedure. Herein, we report a trilayer stent-graft and corresponding delivery system used for the treatment of the AD disease. The stent-graft is made of nitinol stents with an asymmetric Z-wave design and two expanded polytetrafluoroethylene (ePTFE) membranes. Each of the inner and outer surfaces of the stent-graft was covered by an ePTFE membrane, and the two membranes were then sintered together. The biological studies of the sintered ePTFE membranes indicated that the stent-graft had excellent cytocompatibility and hemocompatibility in vitro. Both the stent-graft and the delivery system exhibited satisfactory mechanical properties and operability. The safety and efficacy of this stent-graft and the corresponding delivery system were demonstrated in vivo. In nine canine experiments, the blood vessels of the animals implanted with the stent-grafts were of good patency, and there were no thrombus and obvious stenosis by angiography after implantation for 6 months. Furthermore, all of the nine clinical cases experienced successful implantation using the stent-graft and its postrelease delivery system, and the 1-year follow-ups indicated the preliminary safety and efficacy of the trilayer stent-graft with an asymmetric Z-wave design for interventional treatment.

Funder

National Science Foundation of China

Publisher

Oxford University Press (OUP)

Subject

Biomaterials

Reference80 articles.

1. Acute aortic emergencies—part 2 aortic dissections;White;Adv Emerg Nurs J,2013

2. 2014 ESC guidelines on the diagnosis and treatment of aortic diseases;Erbel;Eur Heart J,2014

3. Early experience and technical aspects of physician-modified fenestration in thoracic endovascular aortic repair for aortic arch pathologies;Li;J Int Med Res,2020

4. Self-radiopaque markers guiding physician-modified fenestration (S-fenestration) in aortic arch endovascular repair;Li;Front Cardiovasc Med,2021

5. Outcomes of hybrid procedure for type B aortic dissection with an aberrant right subclavian artery;Ding;J Vasc Surg,2018

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3