Evaluation of Shimming Techniques on MRI Breast Image Quality at 1.5T

Author:

Zhou Wei1ORCID,Favazza Christopher P1,Axmacher Jessica A1,Trzasko Joshua D1,Geske Jennifer R2,Lee Christine U1

Affiliation:

1. Mayo Clinic, Department of Radiology, Rochester, MN

2. Mayo Clinic, Department of Biomedical Statistics and Informatics, Rochester, MN

Abstract

Abstract Objective The quality of all clinical MRI is dependent on B0 homogeneity, which is optimized during the shimming part of a prescan or preparatory phase before image acquisition. The purpose of this study was to assess shimming techniques clinically employed for breast MRI across our practice, and to determine factors that correlate with higher image quality for contrast-enhanced breast MRI at 1.5T. Methods One hundred consecutive female patients were retrospectively collected with Institutional Review Board approval. Shimming-related parameters, including shim-box placement and shimming gradient offsets were extracted from prior contrast-enhanced 3D fat-suppressed T1-weighted gradient echo image acquisitions. Three breast radiologists evaluated these images for fat saturation, breast density, overall image quality, and artifacts. Technologist experience was also evaluated for variability of shimming. Generalized linear mixed models were used to compare acquisition parameters between fat saturation. P < 0.05 was considered as statistical significance. Results The percentage of soft tissue inside the field of view (FOV) (ie, Tissue/FOV) in the good fat-saturation group (0.37 ± 0.06) was significantly lower (P < 0.01) than that in the poor fat-saturation group (0.39 ± 0.06). Other shimming-related parameters were found not significantly affecting the fat-saturation outcomes. Technologists with more experience tended to have less variable shimming performance than junior technologists did. Conclusions The quality of clinical MRI and especially breast MRI is highly dependent on shimming. Decreasing Tissue/FOV was associated with good image quality (good fat saturation). Optimization of shimming may require manual shimming or higher-order field-correction strategies.

Publisher

Oxford University Press (OUP)

Subject

Radiology, Nuclear Medicine and imaging,Radiological and Ultrasound Technology

Cited by 2 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3