Effects of intraoral aging on mechanical properties of directly printed aligners vs. thermoformed aligners: an in vivo prospective investigation

Author:

Sayahpour Babak1,Zinelis Spiros2,Polychronis Georgios2,Eliades Theodore3ORCID,Goteni Malcolm4,Kopp Stefan1,Eslami Sara1ORCID

Affiliation:

1. Department of Orthodontics, Johann-Wolfgang Goethe University , Theodor-Stern-Kai 7, 60596 Frankfurt am Main , Germany

2. Department of Biomaterials, School of Dentistry, National and Kapodistrian University of Athens , 11527 Athens , Greece

3. Clinic of Orthodontics and Pediatric Dentistry Center of Dental Medicine, University of Zurich , Zurich 8032 , Switzerland

4. Private Orthodontic Practice , Grafenberger Alle 57, 40237 Duesseldorf , Germany

Abstract

Abstract Objective The objective of this study was to examine the impact of intraoral aging on the mechanical properties of directly printed aligners (DPA) compared to thermoformed aligners (TA). Materials and methods This prospective in vivo experiment included three types of aligners: DPAs (group DP) fabricated from Tera Harz TC-85 DAC resin (Graphy, Korea), TA (group INV) made from a polyurethane-based polymer (Align Technology, Inc., CA, USA), and TA (group DUR) made from polyethylene glycol terephthalate based polymer (Scheu-Dental, Germany). Each group was categorized into retrieved (Clin) and unused aligners (Ctr). Thirty patients (10 per group) wore the aligners for 7 days, thereby generating the retrieved samples. Thirty unused aligners were employed as control samples. The following mechanical properties were determined: Martens Hardness (HM), indentation modulus (EIT), elastic index (ηIT), and indentation relaxation (RIT). Intergroup comparisons were conducted using ANOVA/Kruskal–Wallis test. Comparisons between retrieved and control samples were done using Wilcoxon–Mann–Whitney-U/Student’s t-test/Welch’s test. Results Statistically significant differences between the groups were found for both control and used samples (P < .001). Pairwise comparisons also revealed significant differences between the samples. The mechanical properties did not differ significantly between unused and retrieved INV- and DUR-aligners, whereas for DP-aligners significant differences for ηIT and RIT were found following intraoral service (P-values .012 and .002, respectively). Conclusions Group DUR showed generally more favorable mechanical properties compared to DP and INV. The much higher RIT and EIT in DP aligners suggest their higher rigidity and force decay, which could compromise their clinical efficacy.

Publisher

Oxford University Press (OUP)

Subject

Orthodontics

Reference35 articles.

1. Mechanical and geometric properties of thermoformed and 3D printed clear dental aligners;Jindal,2019

2. Advances in orthodontic clear aligner materials;Bichu;Bioact Mater,2023

3. Comparison of dimensional accuracy between direct-printed and thermoformed aligners;Koenig,2022

4. Thermo-mechanical properties of 3D printed photocurable shape memory resin for clear aligners;Lee;Sci Rep-Uk,2022

Cited by 1 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

1. Mechanical and biological properties of current materials of clear aligner;The Journal of The Korean Dental Association;2024-05-31

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3