Comparative Finite Element Modeling Study of Anterior Cervical Arthrodesis Versus Cervical Arthroplasty With Bryan Disc or Prodisc C

Author:

Choi Hoon1,Purushothaman Yuvaraj12,Baisden Jamie L1,Rajasekaran Deepak1,Jebaseelan Davidson2,Yoganandan Narayan1

Affiliation:

1. Center for NeuroTrauma Research, Department of Neurosurgery, Medical College of Wisconsin, Milwaukee, WI 53226, USA

2. School of Mechanical Engineering, Vellore Institute of Technology, Chennai Campus, Chennai, Tamil Nadu 600127, India

Abstract

ABSTRACT Introduction Cervical disc arthroplasty (CDA), a motion-preserving alternative to anterior cervical discectomy and fusion (ACDF), is used in military patients for the treatment of disorders such as spondylosis. Since 2007, the FDA has approved eight artificial discs. The objective of this study is to compare the biomechanics after ACDF and CDA with two FDA-approved devices of differing designs under head and head supported mass loadings. Materials and Methods A previously validated osteoligamentous C2-T1 finite element model was used to simulate ACDF and two types of CDA (Bryan and Prodisc C) at the C5-C6 level. The hybrid loading protocol associated with in vivo head and head supported mass was used to apply flexion and extension loading. First, intact spine was subjected to 2 Nm of flexion extension and the range of motion (ROM) was measured. Next, for each surgical option, flexion-extension moments duplicating the same ROM as the intact spine were determined. Under these surgery-specific moments, ROM and facet force were obtained at the index level, and ROM, facet force, and intradiscal pressure at the rostral and caudal adjacent levels. Results ACDF led to increased motion, force and pressures at the adjacent levels. Prodisc C led to increased motion and facet force at the index level, and decreased motion, facet force, and intradiscal pressure at both adjacent levels. Bryan produced less dramatic biomechanical alterations compared with ACDF and Prodisc C. Numerical results are given in the article. Conclusions Recognizing that ROM is a clinical measure of spine stability/performance, CDA demonstrates a more physiological biomechanical response than ACDF, although the exact pattern depends on the implant design. Anterior and posterior column load-sharing patterns were different between the two implants and may affect implant selection based on the anatomical and pathological state at the index and adjacent levels.

Funder

Office of the Assistant Secretary of Defense for Health Affairs.

Publisher

Oxford University Press (OUP)

Subject

Public Health, Environmental and Occupational Health,General Medicine

Reference35 articles.

1. A review of complication rates for anterior cervical diskectomy and fusion (ACDF);Epstein;Surg Neurol Int,2019

2. Anterior techniques in managing cervical disc disease;Kim;Cureus,2018

3. Performance of cervical arthroplasty at a pseudarthrosed level of a multilevel anterior cervical discectomy and fusion: case report;Spinelli;J Mil Med,2016

4. Single-level Bryan cervical disc arthroplasty: evaluation of radiological and clinical outcomes after 18 years of follow-up;Genitiempo;Eur Spine J

5. Medium-term outcomes of artificial disc replacement for severe cervical disc narrowing;Yeh;J Acute Disease,2014