Comparison of the Effects of Motion and Environment Conditions on Accuracy of Handheld and Finger-Based Pulse Oximeters

Author:

Mark Noy1,Lyubin Anat1,Gerasi Refael1,Ofir Dror2,Tsur Avishai M1,Chen Jacob1,Bader Tarif13

Affiliation:

1. Surgeon General’s Headquarters, Israel Defense Forces Medical Corps, Military POB 02149, Tel Hashomer, Ramat Gan, Israel, Military Postal Code 01215

2. Israel Naval Medical Institute, Israel Defense Forces Medical Corps, Box 22, Rambam Health Care Campus, POB 9602, Haifa, Israel, Postal Code 3109601

3. Faculty of Medicine, Institute for Research in Military Medicine, The Hebrew University of Jerusalem and Israel Defense Forces Medical Corps, POB 12272, Jerusalem, Israel, Postal Code 9112102

Abstract

ABSTRACT Introduction The most common cause of preventable death on the battlefield is significant blood loss, eventually causing decrease in tissue oxygen delivery. Pulse oximeters (POs) are widely used by the Israeli Defense Forces to obtain fast and noninvasive information about peripheral oxygen saturation (SpO2). However, POs are produced by different manufacturers and therefore include different sensors and are based on distinctive algorithms. This makes them susceptible to different errors caused by factors varying from environmental conditions to the severity of injury. The objectives of this study were to compare the reliability of different devices and their accuracy under various conditions. Material and Methods Six POs underwent performance analysis. The finger-based category included: MightySat by Masimo, Onyx II by Nonin, and CMS50D by Contec. The handheld category comprised: RAD5 by Masimo, 9847 model by Nonin, and 3301 model by BCI. Several environmental and physiological parameters were altered using the ProSim8 simulator by Fluke biomedical, forming unique test cases under which the devices were tested in stationary and motion conditions. Results All finger-based POs showed higher error rates of PO SpO2 and heart rate measurements in motion conditions, regardless of the manufacturer. However, newer devices in the handheld category were not affected. Results presented in Phase II showed that the SpO2 measurement error in all the devices was affected by pigmentation. However, the CMS50D, considered a low-cost device, had a significantly higher error size than other devices. In the devices that were influenced both by pigmentation and the finger cleanliness factors, the combined detected error size was clinically significant. The pigmentation, ambient light, and finger cleanliness also had a significant effect on the heart rate measurement in the CMS50D model, unlike the handheld devices, which were not affected. During Phase II, neither the Nonin nor the Masimo devices were deemed to have a significant advantage. Conclusion Considering measurement limitations of POs used is extremely important. Use of handheld devices should be favored for use in motion conditions. Technologically advanced and/or recently developed devices should be preferred because of evolving algorithms, which decrease or eliminate the error factors. The “dirty finger” effect on the measurement error cannot be neglected and therefore the action of finger cleaning should be considered part of the treatment protocol.

Publisher

Oxford University Press (OUP)

Subject

Public Health, Environmental and Occupational Health,General Medicine

Cited by 1 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3