Measurement of Monoclonal Immunoglobulin Protein Concentration in Serum Protein Electrophoresis: Comparison of Automated vs Manual/Human Readings

Author:

Clavijo Alex1,Ryan Nathan1,Xu Hongyan2,Singh Gurmukh13ORCID

Affiliation:

1. Department of Pathology, Medical College of Georgia at Augusta University, Augusta, GA

2. Department of Epidemiology and Bio-Statistics, Medical College of Georgia at Augusta University, Augusta, GA

3. Division of Clinical Pathology, Medical College of Georgia at Augusta University, Augusta, GA

Abstract

Abstract Background Protein concentration of monoclonal immunoglobulin in plasma-cell myeloma/multiple myeloma provides an estimate of the tumor mass and allows for monitoring of the response to treatment. Accurate and reproducible estimates of the monoclonal immunoglobulin concentration are important for patient care. Objective To address the optimum method for estimation of the concentration of monoclonal immunoglobulins. Methods Serum protein electrophoresis and immunofixation electrophoresis were conducted by using the Helena SPIFE Touch instrument. Estimation of the protein concentration of monoclonal immunoglobulin in the gamma region by computer-assisted reading was compared with the reading by technologists and pathology residents, in 300 gels. The data were compared using t-testing and analysis of variance. Results Computer-generated readings had a consistent positive bias. The correlation coefficient of the average reading by technologists and residents with the computer generated value was 0.997. The average positive bias by the computer reading was 0.29 g per dL. The intercept on the regression analysis was 0.22 g per dL. The reading by the computer was significantly higher than each of the human-interpreted readings. The readings by the 3 human groups were not significantly different amongst them. The main reason for the higher reading by the computer was inclusion of a greater area on the anodal size of the peak on the densitometric scan. Conclusions Human- and computer-interpreted readings of the protein concentration of monoclonal immunoglobulin have a high degree of correlation. The consistent positive bias by the computer reading occurred due to inclusion of a greater area of the densitometric scan on the anodal side of the peak. We suggest that vendors should adjust such computer programs to provide readings comparable to those generated by expert humans. We recommend manual delineation of the monoclonal peaks for measuring the concentration of monoclonal immunoglobulins.

Publisher

Oxford University Press (OUP)

Subject

Biochemistry (medical),Clinical Biochemistry

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3