Automated Estimation of Patient’s Size Using Autowed Tool and Indosect Program: A Dosimetric Study for Paediatric Head CT Examinations

Author:

Saeed Mohammed K1,Alshamrani Hassan A1ORCID,Sulieman Abdelmoneim2,Abdullah Youssef M3ORCID,Yousef Mohamed4,Ali Ali S A M5,Alshamrani Khalaf A1

Affiliation:

1. Faculty of Applied Medical Sciences, Najran University , Najran , Saudi Arabia

2. Prince Sattam bin Abdulaziz University , Alkharj , Saudi Arabia

3. Applied Sciences College, Al-Majmaah University , Riyadh , Saudi Arabia

4. Batterjee Medical College , Jeddah , Saudi Arabia

5. Faculty of Science and Technology, Al Neelain University , Khartoum , Sudan

Abstract

Abstract Size-specific dose estimate (SSDE), which can be calculated by measuring the effective diameter (De) or water equivalent diameter (Dw) of the patient, is one of the recent approaches for verifying the individual doses during computer tomography (CT) examinations. This work aimed to compare the Dw estimated by the AutoWED tool and IndoseCT software and to investigate CT axial (ARH) and paediatric head (PH) protocols used in southern Saudi Arabia to calculate the dose received by paediatric patients using metrics of volume CT dose index (CTDIvol) and SSDE. The distribution between the ARH and PH protocols was 57.8 and 42.2%, respectively. There was no significant difference in Dw values between the AutoWED tool and the IndoseCT program (0.13%). Including CT table or other objects during estimation of Dw can lead to variation up to 11.4%. The impact of selecting IndoseCT options to identify the border of the patient may be part of the explanation for these variations. A strong linear relationship was obtained between De and Dw in paediatric head size (R2 = 0.96). Using IndoseCT, for 0–1.5, 1.5–5 and 5.0–18 age groups (years), the Dw was found to be 13.2, 15.3 and 16.8 cm, respectively. The SSDE for the PH protocol was substantially lower than that of the ARH protocol. As a result, education of the individuals engaging in paediatric CT examinations is necessary for dose optimization.

Funder

Deanship of Scientific Research at Najran University

Publisher

Oxford University Press (OUP)

Subject

Public Health, Environmental and Occupational Health,Radiology, Nuclear Medicine and imaging,General Medicine,Radiation,Radiological and Ultrasound Technology

Cited by 2 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

1. Comparison between manual-calculated and IndoseCT-calculated SSDE based on Deff and Dw methods on truncated CT images;Radiation Physics and Chemistry;2024-06

2. Accuracy of IndoseCT for Measuring Effective Diameter from axial CT Images;International Journal of Scientific Research in Science and Technology;2023-02-06