EURADOS REM-COUNTER INTERCOMPARISON AT MAASTRO PROTON THERAPY CENTRE: COMPARISON WITH LITERATURE DATA

Author:

Zorloni Gabriele1ORCID,Bosmans Geert23,Brall Thomas4,Caresana Marco5ORCID,De Saint-Hubert Marijke6,Domingo Carles7,Ferrante Christian5,Ferrulli Francesca18,Kopec Renata9,Leidner Johannes1,Mares Vladimir4,Nabha Racell6,Olko Pawel9ORCID,Caballero-Pacheco Miguel Angel7,Rühm Werner4ORCID,Silari Marco1,Stolarczyk Liliana910ORCID,Swakon Jan9ORCID,Tisi Marco4,Trinkl Sebastian11,Van Hoey Olivier6,Vilches-Freixas Gloria23

Affiliation:

1. CERN , 1211 Geneva 23 , Switzerland

2. Department of Radiation Oncology (MAASTRO) , GROW – School for Oncology, , Maastricht , The Netherlands

3. Maastricht University Medical Centre , GROW – School for Oncology, , Maastricht , The Netherlands

4. Helmholtz Zentrum München, Institute of Radiation Medicine , Ingolstädter Landstr. 1, 85764 Neuherberg , Germany

5. Department of Energy, Polytechnic of Milan , via Lambruschini 4, 20156 Milan , Italy

6. Belgian Nuclear Research Center SCK CEN , 2400 Mol , Belgium

7. Departament de Física, Universitat Autònoma de Barcelona , E-08193 Bellaterra , Spain

8. University of Caen Normandy , 14032 Caen-5 , France

9. Institute of Nuclear Physics PAN , Radzikowskiego 152, 31-342 Krakow , Poland

10. The Danish Centre for Particle Therapy, Aarhus University Hospital , Palle Juul-Jensens Boulevard 25, DK-8200 Aarhus , Denmark

11. Federal Office for Radiation Protection , Medical and Occupational Radiation Protection, Ingolstädter Landstraße 1, 85764 Neuherberg , Germany

Abstract

Abstract The Maastro Proton Therapy Centre is the first European facility housing the Mevion S250i Hyperscan synchrocyclotron. The proximity of the accelerator to the patient, the presence of an active pencil beam delivery system downstream of a passive energy degrader and the pulsed structure of the beam make the Mevion stray neutron field unique amongst proton therapy facilities. This paper reviews the results of a rem-counter intercomparison experiment promoted by the European Radiation Dosimetry Group at Maastro and compares them with those at other proton therapy facilities. The Maastro neutron H*(10) in the room (100–200 μSv/Gy at about 2 m from the isocentre) is in line with accelerators using purely passive or wobbling beam delivery modalities, even though Maastro shows a dose gradient peaked near the accelerator. Unlike synchrotron- and cyclotron-based facilities, the pulsed beam at Maastro requires the employment of rem-counters specifically designed to withstand pulsed neutron fields.

Publisher

Oxford University Press (OUP)

Subject

Public Health, Environmental and Occupational Health,Radiology, Nuclear Medicine and imaging,General Medicine,Radiation,Radiological and Ultrasound Technology

Cited by 3 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献