Evaluation of a cuffless watch-like sensor for 24-hour ambulatory blood pressure monitoring

Author:

Rexhaj E1,Proenca M2,Ambuehl J1,Bonnier G2,Lemay M2

Affiliation:

1. Bern University Hospital, Inselspital, Departments of Clinical Research and Cardiology, Bern, Switzerland

2. CSEM (Swiss Center for Electronics and Microtechnology), Systems Division, Neuchâtel, Switzerland

Abstract

Abstract Introduction Ambulatory blood pressure monitoring (ABPM) is increasingly used in clinical practice for the formal diagnosis of hypertension, and particularly indicated in cases of suspected white-coat effect, masked, or nocturnal hypertension. However, the use of cuffs for ABPM may be painful and cause discomfort, particularly at night, where it may even provoke arousal from sleep and lead to non-representative nighttime blood pressure (BP) values. Purpose To investigate the feasibility of using a cuffless watch-like photoplethysmographic (PPG) sensor for 24-hour ABPM by comparing the PPG-based BP estimates with conventional cuff-derived ABPM values. Methods Our study was approved by the local ethical committee and conducted in 70 participants (43±18 y, 35 with hypertension, 41 male) undergoing cuff-based ABPM. At the contralateral side of the cuff, a cuffless watch-like PPG sensor was worn at the wrist or upper arm. Systolic (SBP) and diastolic (DBP) BP values were estimated by pulse wave analysis on the measured PPG signals. Following a calibration procedure, the PPG-based daytime and nighttime BP estimates were compared to their cuff-based counterparts. The agreement between both methods was evaluated via the mean (bias) and standard deviation (SD) of their differences by Bland-Altman analysis. The agreement on the nocturnal dipping estimates of both devices was also assessed. Finally, the concordance rate (CR) was assessed as the percentage of dipping values showing a concordant direction (dipping vs. non-dipping) between both methods. Results The data of 4 participants were incomplete due to technical issues and had to be rejected prior to analysis. In 4 additional participants, the PPG data quality was insufficient to provide enough BP estimates, probably due to poor sensor tightening. In the remaining 62 participants, we found (see Figure 1) differences between the daytime PPG-based and cuff-based BP estimates of −0.9±3.6 mmHg and −1.4±2.9 mmHg for SBP and DBP, respectively. The differences between the nighttime estimates were −0.8±6.8 mmHg and 0.5±5.3 mmHg, resulting in dipping differences of 0.1±6.8% and −2.0±8.6% for SBP and DBP, respectively. CR on dipping was 97% for both SBP and DBP. Conclusions Good agreement was found between the PPG-based and the cuff-based daytime and nighttime BP averages, with generally negligible (∼1 mmHg) biases. The direction of dipping was highly concordant between both methods. The estimation of its amplitude showed a low bias (∼1%) but a non-negligible spread (SD), which can be in part attributed to the uncertainty on the cuff-based dipping estimates (95% confidence interval range of 12.5% and 16.5% on average for SBP and DBP, respectively), more than twice as large than their PPG-based counterparts (5.7% and 7.8%). Although our study was designed as a method-comparison feasibility study, these results encouragingly suggest that cuffless ABPM may soon become a clinical possibility. Funding Acknowledgement Type of funding sources: None. Figure 1

Publisher

Oxford University Press (OUP)

Subject

Cardiology and Cardiovascular Medicine

Cited by 1 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3