Affiliation:
1. Pfizer Inc., Quality Control, Eastern Point Rd, Groton, CT 06340
Abstract
Abstract
An HPLC procedure for determining sorbitol in bulk sorbitol is described. An Aminex HPX-87 column (Bio-Rad) is used with water as a mobile phase and with a refractive index detector to monitor column eluate. Sorbitol is separated from pentaerythritol, erythritol, ribitol, ethylene glycol, propylene glycol, arabitol, galactitol, mannitol, iditol, and other carbohydrates. The precision of the sorbitol determination, characterized by a 95% confidence interval, corresponds to ±1.22%.
Publisher
Oxford University Press (OUP)
Cited by
1 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献