Affiliation:
1. Department of Cardiac and Thoracic Vascular Surgery, University Medical Center Schleswig-Holstein , Lübeck, Germany
Abstract
Abstract
OBJECTIVES
Selection of a surgical aortic valve (SAV) bioprosthesis model for the treatment of aortic valve disease remains controversial. The aim of this study was to characterize the functional performance of 8 SAV models in a standardized in vitro setting.
METHODS
The hydrodynamic performance of 8 SAVs with labelled size 21 mm (Avalus™, Hancock® II, Mosaic® Ultra™, Perimount®, Perimount® Magna Ease, Epic™ Supra, Trifecta™ GT; Freestyle®), was investigated in a pulse duplicator. Transvalvular pressure gradients and effective orifice area (EOA) were recorded. The geometrical orifice area and physical dimensions of the valves were determined, and new functional dimensions were introduced.
RESULTS
Mean pressure gradient (MPG) and EOA differed significantly between the analysed SAVs. The Epic presented with the lowest EOA and highest MPG, while the Trifecta showed the highest EOA and the lowest MPG. We introduce a useful way to determine the minimal internal diameter and a new measure termed ‘relative orifice area’ to characterize a valve’s performance.
CONCLUSIONS
SAVs showed significant differences in their hydrodynamic performance despite the same label size. This finding was related to the construction of the valves. We introduce a new measure that characterizes the functional performance of a valve model and size for the treatment of an aortic annulus of a specific size. Our data emphasize that SAV selection should carefully be done using an individual patient approach and that future research is necessary to improve the current generation of SAVs.
Publisher
Oxford University Press (OUP)
Subject
Cardiology and Cardiovascular Medicine,Pulmonary and Respiratory Medicine,General Medicine,Surgery
Cited by
4 articles.