Comparative Analyses of Three Point-of-Care Urine Drug Test Devices’ Performance Characteristics for Use in Ambulatory Clinic Settings

Author:

Waggoner Derek C12ORCID,Delaney Sarah R34,Korpi-Steiner Nichole L1

Affiliation:

1. Department of Pathology and Laboratory Medicine, University of North Carolina at Chapel Hill, Chapel Hill NC, USA

2. Department of Pathology and Laboratory Medicine, University of Miami Miller School of Medicine, Miami FL, USA

3. Department of Laboratory Medicine and Pathology, Mayo Clinic, Rochester MN, USA

4. Department of Laboratory Medicine, St. Michael’s Hospital—Unity Health Toronto, Toronto, Ontario Canada

Abstract

Abstract Background Urine drug testing (UDT) is a standard practice used for monitoring controlled and illicit substances in ambulatory care patients. Point-of-care (POC) UDTs are useful tools that allow for drug identification within minutes, providing rapid and objective diagnostic assistance for clinicians. The objective of this study was to evaluate the performance characteristics of 3 different POC UDT devices compared to reference methods. Methods A total of 106 residual urine specimens were collected to evaluate the 3 POC UDT devices: the Profile®-V MEDTOX Scan® drugs of abuse test, Quidel Triage® TOX Drug screen, and Quidel Triage Rapid OXY-BUP-MDMA panel. Device performance was assessed by their ability to identify drug classes/compounds compared to manufacturer and reference method (mass spectrometry) cutoffs. Results The results from quantitative mass spectrometry showed that 77% (84/106) of the samples were positive for one or more drugs. Each device had variable performance across each drug class. Overall, the specificity of the Profile-V MEDTOX Scan test was 90.1%, while the Quidel Triage TOX Drug Screen and Rapid OXY-BUP-MDMA devices had specificities of 89.0% and 50.0% using their respective manufacturer-stated cutoffs. Overall sensitivity was determined to be 98.6%, 97.0%, and 100% for the Profile-V MEDTOX Scan, Quidel Triage TOX Drug Screen, and Rapid OXY-BUP-MDMA, respectively. Conclusions Of the 3 POC UDT devices evaluated, the Profile-V MEDTOX Scan demonstrated the best overall sensitivity and specificity compared to reference methods. False positive and negative results are possible with UDTs, ultimately the best device may depend on patient population and drugs of interest.

Publisher

Oxford University Press (OUP)

Subject

General Medicine

Cited by 1 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3