Multicenter Evaluation of a New, Fully Automated Androstenedione Electrochemiluminescence Immunoassay: Precision Analysis, Method Comparison, and Determination of Reference Ranges

Author:

Obermayer-Pietsch Barbara1,de Ramon Marta2,Reichmuth Claudia3,Bendig Garnet4,Hutzler Stefan4,Taibon Judith4,Rank Christopher M4,Findeisen Peter5

Affiliation:

1. Department of Internal Medicine, Division of Endocrinology and Diabetology, Endocrinology Lab Platform, Medical University of Graz, Graz, Austria

2. Clinical Biochemistry, Hormones and Tumor Markers Section, Laboratori de Referència de Catalunya, Barcelona, Spain

3. Agent from TRIGA-S representing Roche Diagnostics GmbH, Penzberg, Germany

4. Roche Diagnostics GmbH, Penzberg, Germany

5. Study Department, MVZ Labor Dr. Limbach & Kollegen, Heidelberg, Germany

Abstract

Abstract Background Androstenedione (ASD) levels can aid diagnosis of hyperandrogenism together with other clinical/laboratory findings. We evaluated performance of the new, automated Elecsys® ASD assay vs an ASD isotope dilution-liquid chromatography-tandem mass spectrometry (ID-LC-MS/MS) reference measurement procedure and determined reference ranges. Methods Repeatability/intermediate precision were assessed using 3 control levels and 5 human serum pools (n = 75 each; Clinical and Laboratory Standards Institute EP05-A3). Method comparisons vs commercially available immunoassays [IMMULITE ASD (Siemens) and LIAISON ASD (DiaSorin)] and an ID-LC-MS/MS measurement procedure method were conducted using 421 serum samples; Passing–Bablok regression and Pearson’s correlation coefficients were calculated. Reference ranges and distribution of values associated with polycystic ovary syndrome (PCOS) were determined in five clinical cohorts using samples from several sites/vendors. Results Repeatability/intermediate precision coefficients of variation across all sites were 2.01% to 3.91% and 2.43% to 4.30%, respectively (mean ASD: 7.80–34.7 nmol/L). The Elecsys ASD assay showed poor agreement with IMMULITE ASD (slope = 0.459; r = 0.856; n = 320), fair agreement with LIAISON ASD (slope = 0.625; r = 0.984; n = 327), and very good agreement with ID-LC-MS/MS (slope = 1.040; r = 0.996; n = 332). Reference ranges (2.5th–97.5th percentiles) were: children (≤8 years; n = 140), <0.525 to 1.81 nmol/L; males (≥18 years; n = 138), 0.979 to 5.32 nmol/L; and postmenopausal females (n = 140), 0.654 to 3.74 nmol/L. Reference range (5th–95th percentiles) for females with fertile cycle (≥18 years; n = 84) was 1.71 to 4.58 nmol/L. The distribution of values (2.5th–97.5th percentiles) in females with PCOS (n = 125) was 2.26 to 12.1 nmol/L. Conclusions Elecsys ASD assay demonstrated excellent precision and very good agreement with ID-LC-MS/MS. Reference ranges were established to support results interpretation in routine practice.

Publisher

Oxford University Press (OUP)

Subject

General Medicine

Cited by 2 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献