Recent Recall of Iron Reagent—Investigation of Potential Reagent Contamination and Assay Improvement Strategy

Author:

Narasimhan Madhusudhanan1,Jaleta Kefyalew2,Adhikari Shishir2,Berihun Mizanu2,SataraNatarajan Kavithalakshmi1ORCID,Mahimainathan Lenin1,Cao Jing13,Jones Patricia Mary13ORCID,Hashim Ibrahim14,Muthukumar Alagar R1

Affiliation:

1. Department of Pathology, University of Texas Southwestern Medical Center , Dallas, TX , United States

2. Clements University Hospital, University of Texas Southwestern Medical Center Clinical Laboratory Services, , Dallas, TX , United States

3. Main Laboratory, Children’s Medical Center , Dallas, TX , United States

4. Department of Pathology, Parkland Hospital System , Dallas, TX , United States

Abstract

Abstract Background Recently, a major manufacturer recalled several lots of iron assay reagent due to positive bias of roughly 15%–30% and the cause remains unknown. This study investigated the root cause of this positive bias and evaluated a simple practical approach to improve the assay. Methods Performance comparison of recalled and unimpacted iron assay kits was done utilizing calibrators, quality control (QC) materials, and 42 remnant patient samples. Spectral scan and trace elements analysis of R1 and R2 reagents was performed. Copper (Cu) and thiourea (TU) spiking experiments were utilized to elucidate the cause and prevention of positive bias seen with recalled lots. Results Iron measurements in QC materials and patient samples using recalled reagents generated a positive bias of 17.5% and 21%, respectively. Correspondingly, the recalled R2 reagents, but not R1, showed a rise in basal absorbance along with an unanticipated presence of Cu (22.7 µg/dL) and lead (7.5 µg/L). Cu spiking to recalled and unimpacted R2 reagent intensified the reagent color besides falsely increasing its absorbance, calibration factor, and patient iron measurements. Interestingly, addition of TU (65 mmol/L) to R2 reagent from unimpacted lot prevented the short-term and prolonged Cu-induced spurious rise in calibration factor and patient iron estimations. Conclusions We conclude that accidental copper contamination of R2 reagent during manufacturing could be a reason underlying the positive bias in the recalled iron reagent lots. Addition of TU in ferene-containing R2 reagent is a simple and effective means to prevent Cu-induced false elevation in iron values.

Publisher

Oxford University Press (OUP)

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3