Performance of a Norfentanyl Immunoassay in Specimens with Low Concentrations of Fentanyl and/or Norfentanyl

Author:

Uljon Sacha12ORCID,Tolan Nicole V23,Mahowald Grace K12ORCID,Khaliq Tahira1,Urwiller Elizabeth D3,Fernandes Maria Daluz3,Basu Sankha S23,Kang Phillip3,Erickson Timothy B24,Hayes Bryan D25,Chai Peter R24,Melanson Stacy E F23

Affiliation:

1. Department of Pathology, Massachusetts General Hospital , Boston, MA , United States

2. Department of Pathology, Harvard Medical School , Boston, MA , United States

3. Department of Pathology, Brigham and Women’s Hospital , Boston, MA , United States

4. Department of Emergency Medicine, Division of Medical Toxicology, Brigham and Women’s Hospital , Boston, MA , United States

5. Department of Pharmacy, Massachusetts General Hospital , Boston, MA , United States

Abstract

Abstract Background Many fentanyl immunoassays are limited in their ability to detect norfentanyl. Urine specimens collected from individuals who have been exposed to fentanyl frequently have detectable concentrations of norfentanyl (≥2 ng/mL) but low concentrations of fentanyl (<2 ng/mL) by LC-MS/MS. The Lin-Zhi Fentanyl II Immunoassay (Lin-Zhi) claims 100% cross-reactivity with norfentanyl and therefore may detect exposure missed by other assays. Methods In addition to verifying the manufacturer's analytical sensitivity claims, we selected 92 urine specimens with low-positive Lin-Zhi results (1–99 absorbance units, lowest 10%) for analysis by the Immunalysis Health Equity Impact Assessment and ARK II fentanyl methods. The accuracy of the 3 immunoassays was compared to LC-MS/MS as the reference method. Results Spiking studies using purified fentanyl and norfentanyl and a set of 100 consecutive specimens confirmed the manufacturer's claims of limit of detection for fentanyl (3.8 ng/mL) and norfentanyl (5.0 ng/mL). However, the 92 low-positive patient specimens demonstrated concentrations of norfentanyl and fentanyl below 2.0 ng/mL by LC-MS/MS, with 47 (51%) having only norfentanyl detected. When comparing Lin-Zhi to the Immunalysis and ARK II immunoassays, only 27 (29%) of the 92 specimens were concordant. Fifty-two (57%) of the specimens were positive by LC-MS/MS and Lin-Zhi but false negative by one or both other immunoassays. Seven specimens (8%) were positive by Lin-Zhi but negative by the other immunoassays and had undetectable concentrations (<2 ng/mL) of fentanyl and norfentanyl by LC-MS/MS. Conclusions The clinical sensitivity of the Lin-Zhi exceeds the manufacturer’s claims, providing results comparable to LC-MS/MS methods.

Publisher

Oxford University Press (OUP)

Cited by 1 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

1. Choosing a Fentanyl Immunoassay;The Journal of Applied Laboratory Medicine;2024-07-13

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3