Quantifying the dosimetric effects of neck contour changes and setup errors on the spinal cord in patients with nasopharyngeal carcinoma: establishing a rapid estimation method

Author:

Li Yinghui12,Wei Zhanfu3,Liu Zhibin2,Teng Jianjian2,Chang Yuanzhi2,Xie Qiuying2,Zhang Liwen2,Shi Jinping2,Chen Lixin1

Affiliation:

1. State Key Laboratory of Oncology in South China, Sun Yat-sen University Cancer Center, Sun Yat-Sen University of Medical Sciences, Guangzhou, 510060, Guangdong, China

2. Physics Department of the Radiotherapy Department, The First People’s Hospital of FoShan (Affiliated FoShan Hospital of Sun Yat-sen University), Foshan, 528000, Guangdong, China

3. Radiotherapy Center of the Oncology Medical Center, The First People’s Hospital of ZhaoQing, Zhaoqing, 526000, Guangdong, China

Abstract

Abstract The purpose of this study was to quantify the effect of neck contour changes and setup errors on spinal cord (SC) doses during the treatment of nasopharyngeal carcinoma (NPC) and to establish a rapid dose estimation method. The setup errors and contour changes in 60 cone-beam computed tomography (CBCT) images of 10 NPC patients were analysed in different regions of the neck (C1–C3, C4–C5 and C6–C7). The actual delivered dose to the SC was calculated using the CBCT images, and univariate simulations were performed using the planning CT to evaluate the dose effects of each factor, and an index ${\mathrm{Dmax}}_{\mathrm{displaced}}$ was introduced to estimate the SC dose. Compared with the planned dose, the mean (maximum) Dmax increases in the C1–C3, C4–C5 and C6–C7 regions of the SC were 2.1% (12.3%), 1.8% (8.2%) and 2.5% (9.2%), respectively. The simulation results showed that the effects of setup error in the C1–C3, C4–C5 and C6–C7 regions were 1.5% (9.7%), 0.9% (8.2%) and 1.3% (6.3%), respectively, and the effects of contour change were 0.4% (1.7%), 0.7% (2.5%) and 1.5% (4.9%), respectively. The linear regression model can be used to estimate the dose effect of contour changes (R2 > 0.975) and setup errors (R2 = 0.989). Setup errors may lead to a significant increase in the SC dose in some patients. This study established a rapid dose estimation method, which is of great significance for the daily dose evaluation and the adaptive re-planning trigger of the SC.

Funder

Medical Science and Technology Foundation of Guangdong Province

Spacial fund of Foshan Summit plan

Publisher

Oxford University Press (OUP)

Subject

Health, Toxicology and Mutagenesis,Radiology, Nuclear Medicine and imaging,Radiation

Cited by 3 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3