Tolerance levels of mass density for adaptive helical tomotherapy using MVCT

Author:

Tsunemine Shogo123,Ozawa Shuichi3456,Nakao Minoru456,Miura Hideharu756,Saito Akito56,Kawahara Daisuke56,Onishi Yasuhiko3,Onishi Takashi3,Okawa Fumito3,Terai Atsushi3,Hashiguchi Taiki3,Yamasaki Hidetoshi3,Maruta Tsutomu8,Murakami Yuji456,Nagata Yasushi456

Affiliation:

1. Graduate School of Biomedical and Health Sciences Program of Medicine Doctoral Course, , , 1-2-3, Kasumi, Minamiku, Hiroshima, 734-8553   Japan

2. Hiroshima University Program of Medicine Doctoral Course, , , 1-2-3, Kasumi, Minamiku, Hiroshima, 734-8553   Japan

3. National Hospital Organization Himeji Medical Center Department of Radiology, , 68, Hommachi, Himeji, Hyogo, 670-8520 , Japan

4. Hiroshima High-Precision Radiotherapy Cancer Center , Hiroshima, 3-2-2, Futabanosato, Higashiku, Hiroshima, 732-0057 , Japan

5. Graduate School of Biomedical and Health Sciences Department of Radiation Oncology, , , 1-2-3, Kasumi, Minamiku, Hiroshima, 734-8553   Japan

6. Hiroshima University Department of Radiation Oncology, , , 1-2-3, Kasumi, Minamiku, Hiroshima, 734-8553   Japan

7. Hiroshima High-Precision Radiotherapy Cancer Center, Hiroshima , 3-2-2, Futabanosato, Higashiku, Hiroshima, 732-0057 , Japan

8. National Hospital Organization Himeji Medical Center Department of Therapeutic Radiology, , 68, Hommachi, Himeji, Hyogo, 670-8520 , Japan

Abstract

Abstract Daily dose distributions for adaptive radiotherapy (ART) using helical tomotherapy (HT) are calculated using megavoltage computed tomography (MVCT). Generally, the MVCT number is converted to mass density (MD) using an MD calibration table (MVCT-MD table). The aims of this study are to calculate the tolerance levels of the MD for ART and to evaluate the tolerance levels using clinical patient plans. These tolerance levels of MD were calculated based on the tissue maximum ratio (TMR) of 6MV flattening-filter-free (FFF) beam of HT and the effective tissue thickness data from an International Commission on Radiological Protection 110 phantom data for lung, adipose/muscle and cartilage/spongy-bone. These tolerance levels were determined by considering both the MD causing a dose error of 2% and the variation in MVCT numbers. Subsequently, the stability of the MD values was estimated with the standard deviations (SD) in the MVCT number over 6 months. The dose distribution for clinical patient plans was calculated using the MVCT-MD table with added tolerance levels. These tolerance levels were determined as MD differences causing a dose error of 2%, and were ± 0.049 g/cm3, ± 0.030 g/cm3 and ± 0.049 g/cm3 for lung, adipose/muscle and cartilage/spongy-bone, respectively. The calculated dose distribution errors using the MVCT-MD table added tolerance levels were within 2%. We proposed these tolerance levels in MD for the quality control of the MVCT-MD table.

Funder

AMED

JSPS

Publisher

Oxford University Press (OUP)

Subject

Health, Toxicology and Mutagenesis,Radiology, Nuclear Medicine and imaging,Radiation

Reference17 articles.

1. History of tomotherapy;Mackie;Phys Med Biol,2006

2. Tomotherapy as a tool in image-guided radiation therapy (IGRT): current clinical experience and outcomes;Yartsev;Biomed Imaging Interv J,2007

3. A technique for adaptive image-guided helical tomotherapy for lung cancer;Ramsey;Int J Radiat Oncol Biol Phys,2006

4. The use of megavoltage CT (MVCT) images for dose recomputations;Langen;Phys Med Biol,2005

5. Improving dose calculations on tomotherapy MVCT images Improving dose calculations on tomotherapy MVCT images;Crop;J Appl Clin Med Phys,2012

Cited by 2 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3