Objectives
To test the performance characteristics of 69 primary immunohistochemistry antibodies after expiration and compare with fresh primary antibodies wherever possible.
Methods
A total of 69 expired primary antibodies were evaluated for specificity, background staining, and intensity. An optimal staining result corresponded to a semiquantitatively scored 2+ or 3+ intensity, with intact specificity devoid of moderate or strong background staining. Any deviation from a normal staining pattern was also considered to be a suboptimal result.
Results
Nearly half of the antibodies studied showed an optimally positive staining result after expiration (34/69, 49.2%). Overall, 10.1% (7/69) of antibodies could be compared with fresh primary antibodies of the same clone with equivalent results. Eight of 69 (11.6%) expired antibodies showed splotchy or granular staining.
Conclusions
Evidence from this study and previous work point to maintained functionality of a fair number of primary immunohistochemical antibodies after expiration. Decisions about the use of such reagents should be guided by a thorough assessment of functionality by the pathologist rather than a manufacturer-specified deadline. Quality maintenance should imply a sensible balance between histopathologic performance and economics.
Publisher
Oxford University Press (OUP)
Reference12 articles.
1. Antibody expiration in the context of resource limitation: what is the evidence basis?;Savage;Am J Clin Pathol.,2010
2. Extension of useful reagent shelf life beyond manufacturers’ recommendations. Cell Markers Committee of the College of American Pathologists;Tubbs;Arch Pathol Lab Med.,1998
3. Satisfactory performance of primary antibodies beyond manufacturers’ recommended expiration dates;Balaton;Appl Immunohistochem Mol Morphol.,1999
Cited by
2 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献