A New Method for Dead Time Calibration and a New Expression for Correction of WDS Intensities for Microanalysis

Author:

Donovan John J1ORCID,Moy Aurélien2,von der Handt Anette3,Gainsforth Zack4ORCID,Maner James L5,Nachlas William2,Fournelle John2ORCID

Affiliation:

1. CAMCOR, University of Oregon , 1443 East 13th Ave, Eugene, OR 97403 , USA

2. Geoscience Department, University of Wisconsin-Madison , 1215 W Dayton St, Madison, WI 53706 , USA

3. EOAS, University of British Columbia , 2207 Main Mall #2020, Vancouver, BC V6T 1Z4 , Canada

4. Space Sciences Laboratory, University of California at Berkeley , 7 Gauss Way, Berkeley, CA 94720 , USA

5. Nuclear Materials Science Group, Materials Science and Technologist Division, Los Alamos National Laboratory , PO BOX 1663, Los Alamos, NM 87545 , USA

Abstract

Abstract Observed photon count rates must be corrected for detector dead time effects for accurate quantification, especially at high count rates. We present the “constant k-ratio” method, a new approach for calibrating dead time for wavelength dispersive spectrometers by measuring k-ratios as a function of beam current. The method is based on the observation that for a given emission line at a specific take-off angle and electron beam energy, the intensity ratio from two materials containing the element should remain constant as a function of beam current, if the dead time calibration is accurate. The method has the advantage that it does not rely on the linearity of the beam current picoammeter, yet also allows the analyst to evaluate the picoammeter linearity, another critical parameter in EPMA calibration. By simultaneously comparing k-ratios for all spectrometers, one can also ascertain k-ratio consensus, essential for inter-laboratory comparisons. We also introduce improved dead time expressions and provide best practices on how to perform these instrument calibrations using this new “constant k-ratio” method. These improvements enable quantitative analysis of major and minor elements with high accuracy at high beam currents, simultaneously with trace elements with high sensitivity, for point analyses and X-ray mapping.

Publisher

Oxford University Press (OUP)

Subject

Instrumentation

Reference13 articles.

1. Real time digital pulse processing for X-ray and gamma ray semiconductor detectors;Abbene;Nucl Instrum Methods Phys Res,2013

2. Electron Beam Microanalysis

3. Accuracy of quantitative electron probe microanalysis with energy dispersive spectrometers;Beaman;Anal Chem,1972

4. X-ray secondary-emission (fluorescence) spectrometry; General introduction;Bertin;Principles Pract X-Ray Spectromet Anal,1975

5. Quantitative WDS compositional mapping using the electron microprobe;Donovan;Am Mineral: J Earth Planetary Mater,2021