Safety of smartwatches and their chargers in patients with cardiac implantable electronic devices

Author:

Tzeis Stylianos1ORCID,Asvestas Dimitrios1,Moraitis Nektarios2,Vardas Emmanuel P1,Mililis Panagiotis3,Letsas Konstantinos3,Kouvelas Konstantinos1,Nikita Konstantina S2,Vardas Panos1

Affiliation:

1. Department of Cardiology, Mitera General Hospital, Hygeia Group, Erithrou Stavrou 6, Maroussi, Athens, Greece

2. School of Electrical and Computer Engineering, National Technical University of Athens, Athens, Greece

3. Department of Cardiology, Electrophysiology Laboratory Evangelismos General Hospital, Athens, Greece

Abstract

AbstractAimsCardiac implantable electronic devices (CIEDs) are susceptible to electromagnetic interference (EMI). Smartwatches and their chargers could be a possible source of EMI. We sought to assess whether the latest generation smartwatches and their chargers interfere with proper CIED function.Methods and resultsWe included consecutive CIED recipients in two centres. We tested two latest generation smartwatches (Apple Watch and Samsung Galaxy Watch) and their charging cables for potential EMI. The testing was performed under continuous electrocardiogram recording and real-time device telemetry, with nominal and ‘worst-case’ settings. In vitro magnetic field measurements were performed to assess the emissions from the tested devices, initially in contact with the probe and then at a distance of 10 cm and 20 cm. In total, 171 patients with CIEDs (71.3% pacemakers–28.7% implantable cardioverter-defibrillators) from five manufacturers were enrolled (63.2% males, 74.8 ± 11.4 years), resulting in 684 EMI tests. No EMI was identified in any patient either under nominal or ‘worst-case scenario’ programming. The peak magnetic flux density emitted by the smartwatches was similar to the background noise level (0.81 μT) even when in contact with the measuring probe. The respective values for the chargers were 4.696 μΤ and 4.299 μΤ for the Samsung and Apple chargers, respectively, which fell at the background noise level when placed at 20 cm and 10 cm, respectively.ConclusionTwo latest generation smartwatches and their chargers resulted in no EMI in CIED recipients. The absence of EMI in conjunction with the extremely low intensity of magnetic fields emitted by these devices support the safety of their use by CIED patients.

Publisher

Oxford University Press (OUP)

Subject

Physiology (medical),Cardiology and Cardiovascular Medicine

Cited by 5 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370