Acute safety, efficacy, and advantages of a novel cryoballoon ablation system for pulmonary vein isolation in patients with paroxysmal atrial fibrillation: initial clinical experience

Author:

Anic Ante1ORCID,Lever Nigel2,Martin Andrew2,Breskovic Toni1,Sulkin Matthew S3,Duffy Elizabeth3,Saliba Walid I4,Niebauer Mark J4,Wazni Oussama M4,Varma Niraj4

Affiliation:

1. Cardiovascular Diseases, University Hospital Center Split, Spinciceva 1, 21000 Split, Croatia

2. Cardiovascular Diseases, Green Lane Cardiovascular Services, Auckland City Hospital, Auckland, New Zealand

3. Electrophysiology, Boston Scientific Corp., St. Paul, MN, USA

4. Department for Cardiovascular Diseases, Cleveland Clinic Foundation, Heart and Vascular Institute, Cleveland, OH, USA

Abstract

Abstract Aims Cryoballoon pulmonary vein isolation (PVI) is a safe and effective treatment for atrial fibrillation (AF). Current limitations include incomplete vein occlusion due to balloon rigidity and inconsistent electrogram recording, which impairs identification of isolation. We aimed to evaluate the acute safety and performance of a novel cryoballoon system. Methods and results The system includes a steerable sheath, mapping catheter, and a balloon that maintains uniform inflation pressure and size following initiation of ablation. Protocol-directed cryoablation was delivered for 180 s for isolation documented in ≤60 s, otherwise freeze duration was 240 s. Primary endpoints were acute safety and vein isolation. Pulmonary vein isolation was confirmed at ≥30 min post-isolation. Data were compared across vein locations. Thirty patients with paroxysmal AF were enrolled at two centres and underwent PVI. Pulmonary vein isolation was achieved with cryoablation only in 100% of veins (120/120). Nadir temperature was −53.1 ± 5.3°C. The number of applications to achieve PVI was 1.4 ± 0.4 per vein. Of the 120 veins, 89 were isolated with a single cryothermal application (10/30 patients required only 4 total cryoablations). There were no procedural- or device-related serious adverse events at 30 days post-procedure. A subset (24/30) of patients was followed for 1-year and 71% (17/24) remained free of atrial arrhythmias. Six patients with arrhythmia recurrence were remapped and three had durable PVI for all four veins. Conclusion In this first human experience, the novel cryoballoon platform was safe, efficacious, and demonstrated a high proportion of successful single ablation isolation.

Funder

Cryterion Medical Inc

Publisher

Oxford University Press (OUP)

Subject

Physiology (medical),Cardiology and Cardiovascular Medicine

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3