Electrical characteristics of pacing different portions of the His bundle in bradycardia patients

Author:

Hu Yiran12,Gu Min1,Hua Wei1,Niu Hongxia1,Li Hui3,Chen Xuhua1,Zhang Nixiao1,Liu Xi1,Zhang Shu1

Affiliation:

1. The Cardiac Arrhythmia Center, State Key Laboratory of Cardiovascular Disease, National Clinical Research Center of Cardiovascular Diseases, Fuwai Hospital, National Center for Cardiovascular Diseases, Chinese Academy of Medical Sciences and Peking Union Medical College, No. 167, Beilishi Road, Xicheng District, Beijing 100037, China

2. Department of Cardiology and Macrovascular Disease, Beijing Tiantan Hospital, Capital Medical University, Beijing, China

3. Department of Ultrasound, Fuwai Hospital, National Center for Cardiovascular Disease, Beijing, China

Abstract

Abstract Aims His-bundle pacing (HBP) can be achieved in either atrial-side HBP (aHBP) or ventricular-side HBP (vHBP). The study compared the pacing parameters and electrophysiological characteristics between aHBP and vHBP in bradycardia patients. Methods and results Fifty patients undergoing HBP implantation assisted by visualization of the tricuspid valvular annulus (TVA) were enrolled. The HBP lead position was identified by TVA angiography. Twenty-five patients were assigned to undergo aHBP and compared with 25 patients who underwent vHBP primarily in a prospective and randomized fashion. Pacing parameters and echocardiography were routinely assessed at implant and 3-month follow-up. His-bundle pacing was successfully performed in 45 patients (90% success rate with 44.4% aHBP and 55.6% vHBP). The capture threshold was lower in vHBP than aHBP at implant (vHBP: 1.1 ± 0.5 vs. aHBP: 1.4 ± 0.4 V/1.0 ms, P = 0.014) and 3-month follow-up (vHBP: 0.8 ± 0.4 vs. aHBP: 1.7 ± 0.8 V/0.4 ms, P < 0.001). The R-wave amplitude was higher in vHBP than in aHBP at implant (vHBP: 4.5 ± 1.4 vs. aHBP: 2.0 ± 0.8 mV, P < 0.001) and at 3-month follow-up (vHBP: 4.4 ± 1.5 vs. aHBP: 1.8 ± 0.7 mV, P < 0.001). No procedure-related complications and aggravation of tricuspid valve regurgitation were observed in most patients and echocardiographic assessment of cardiac function remained in the normal range in all patients during the follow-up. Conclusion This study demonstrates that vHBP features a low and stable pacing capture threshold and high R-wave amplitude, suggesting better pacing mode management and battery longevity can be achieved by HBP in the ventricular side.

Publisher

Oxford University Press (OUP)

Subject

Physiology (medical),Cardiology and Cardiovascular Medicine

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3