Clinical outcomes after implantation of quadripolar compared to bipolar left ventricular leads in patients undergoing cardiac resynchronization therapy: a systematic review and meta-analysis

Author:

Erath Julia W1ORCID,Benz Alexander P1ORCID,Hohnloser Stefan H1,Vamos Mate1

Affiliation:

1. Division of Clinical Electrophysiology, Department of Cardiology, University Hospital Frankfurt, Goethe University, Theodor-Stern-Kai 7, D Frankfurt, Germany

Abstract

Abstract Aims Some retrospective and prospective studies in heart failure patients with indication for cardiac resynchronization therapy (CRT) suggest better clinical outcomes for quadripolar (QP) left ventricular (LV) leads over bipolar (BP) leads. Although, lead failure remains an important safety concern, when using these more complex, novel electrodes. To evaluate safety and efficacy outcomes for QP vs. BP LV leads in patients receiving CRT. Methods and results We performed a comprehensive literature search through 2018 in PubMed, Cochrane Library, and Google Scholar databases to identify studies comparing patients with QP and BP LV CRT leads. A total of 12 studies were selected for analysis comprising 31 403 patients (QP lead: 22 429 patients; BP lead: 8974 patients). Eight studies examined the effects of CRT on survival. In these studies, use of QP electrodes was associated with significantly better survival compared to patients with BP LV leads (OR 0.61, 95% CI 0.50–0.76; P < 0.01). Clinical improval measured in New York Heart Association functional class (OR 0.59, 95% CI 0.34–1.01; P = 0.05) and hospitalization rates (OR 0.67, 95% CI 0.55–0.83; P < 0.01) were also improved in patients receiving QP leads. Lead malfunctions defined as LV lead failure resulting in lead deactivation (OR 0.57, 95% CI 0.34–0.98; P = 0.04) or LV lead dislodgement requiring LV lead replacement/repositioning (OR 0.48; 95% CI 0.31–0.75; P < 0.01) were more often encountered among patients with BP leads compared to patients with QP leads. Conclusion Our meta-analysis suggests distinct benefits of QP over BP electrodes in patients undergoing CRT.

Publisher

Oxford University Press (OUP)

Subject

Physiology (medical),Cardiology and Cardiovascular Medicine

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3