Projected longevities of cardiac implantable defibrillators: a retrospective analysis over the period 2007–17 and the impact of technological factors in determining longevity

Author:

Paton Maria F1ORCID,Landolina Maurizio2,Billuart Jean-Renaud3,Field Duncan4,Sibley Jonathan5,Witte Klaus1

Affiliation:

1. Leeds Institute of Cardiovascular and Metabolic Medicine, University of Leeds, Leeds, UK

2. Department of Cardiology, ASST. Ospedale Maggiore, Crema, Italy

3. MicroPort CRM, Clamart, France

4. Department of Cardiology, Colchester General Hospital, Essex, UK

5. Departemt of Cardiac Physiology, Essex Cardio-thoracic Centre, Basildon, UK

Abstract

Abstract Aims Implanters of cardiac implantable electronic devices cannot easily choose devices by longevity as usually current models only have projected longevity data since those with known performance are obsolete. This study examines how projected device longevities are derived, the influencing factors, and their roles in guiding model choice. Methods and results Ninety-eight implantable cardioverter-defibrillator (ICD) and cardiac resynchronization therapy-defibrillator (CRT-D) models released in Europe in 2007–17 were analysed for reported battery capacities, projected longevities for standardized settings stipulated by the French Haute Autorité de Santé (HAS) and manufacturer-chosen settings. Battery capacities and HAS projected longevities increased during the study period. Based on current drain estimation, therapy functions consumed only a small portion (2–7%) of the battery energy for single- and dual-chamber ICDs, but up to 50% (from biventricular pacing) for CRT-Ds. Large differences exist between manufacturers and models both in terms of battery capacity and energy consumption. Conclusion Battery capacity is not the sole driver of longevity for electronic implantable cardiac devices and, particularly for ICDs, the core function consume a large part of the battery energy even in the absence of therapy. Providing standardized current drain consumption in addition to battery capacity may provide more meaningful longevity information among implantable electronic cardiac devices.

Funder

NIHR

Publisher

Oxford University Press (OUP)

Subject

Physiology (medical),Cardiology and Cardiovascular Medicine

Reference20 articles.

Cited by 7 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献