Abstract
Abstract
To compare radiation dose and image quality of three CT-scanners using optimal dose protocols in patients with suspected urolithiasis regarding additional hardware (tin filter) and software (iterative reconstruction). Examinations from a single-source CT-scanner (A2) and a dual-source CT-scanner (DSCT) (A1) were compared to a tin filter DSCT (B) regarding dose-length product (DLP) and volume-weighted CT dose-index (CTDIvol). DLP of B was 51 and 53% lower in comparison to A1 and A2 (78.62, 159.20 and 165.80 mGy·cm, respectively; P < 0.0001). CTDIvol of B was 53% and 56% significantly lower compared to A1 and A2, respectively (1.52 vs. 3.22 vs. 3.46 mGy; P < 0.0001). Image quality in B proved to be similar to A1 and A2 (3.57, 3.51 and 3.60, respectively; P > 0.05). Inter-rater agreement regarding image quality was good for all CT-scanners (κ = 0.62). Modern CTs with a built-in tin filter allow a significant reduction of radiation exposure in patients with suspected urolithiasis by optimizing the X-ray spectrum.
Funder
German Research Foundation
Publisher
Oxford University Press (OUP)
Subject
Public Health, Environmental and Occupational Health,Radiology, Nuclear Medicine and imaging,General Medicine,Radiation,Radiological and Ultrasound Technology