IMAGE QUALITY STUDY OF RADIATION-REDUCED COMBINED CHEST AND ABDOMEN/PELVIS CT COMPARED WITH A STANDARD PROTOCOL

Author:

Bos Denise1ORCID,Zensen Sebastian1ORCID,Opitz Marcel1,Haubold Johannes1,Forsting Michael1,Nassenstein Kai1,Guberina Nika2,Wetter Axel13

Affiliation:

1. Institute of Diagnostic and Interventional Radiology and Neuroradiology, University Hospital Essen, Hufelandstrasse 55, 45147 Essen, Germany

2. Department of Radiotherapy, West German Cancer Center, University Hospital Essen, Hufelandstrasse 55, 45147 Essen, Germany

3. Department of Diagnostic and Interventional Radiology, Neuroradiology, Asklepios Klinikum Harburg, Eißendorfer Pferdeweg 52, 21075 Hamburg, Germany

Abstract

Abstract The purpose of this study is to examine radiation doses and image quality of a low-dose (LD) protocol for chest and abdomen/pelvis (CAP) CT compared with a standard (STD) protocol. A total of 361 patients were included between October 2019 and April 2020; 104 patients with LD-protocol (100 kV, ref mAs 80 (chest)/145 (abdomen/pelvis)) and 257 patients with STD-protocol (100 kV, ref mAs 100 (chest)/180 (abdomen/pelvis)) at second-generation dual-source CT. Radiation doses for CTDIvol and DLP, and objective and subjective image qualities of 50 examinations from each group were evaluated. The LD-protocol applied significantly lower radiation doses compared with the STD-protocol (p < 0.001), achieving a dose reduction by 37% for the median DLP in chest, 19% in abdomen/pelvis and 22% in total. Median total DLP was 342 mGy·cm (LD) vs. 436 mGy·cm (STD). The LD-CAP CT protocol achieved a significant dose reduction far below national diagnostic reference levels, ensuring acceptable and good image quality.

Funder

University Medicine Essen Academy

German Research Foundation

Faculty of Medicine, University of Duisburg-Essen

Publisher

Oxford University Press (OUP)

Subject

Public Health, Environmental and Occupational Health,Radiology, Nuclear Medicine and imaging,General Medicine,Radiation,Radiological and Ultrasound Technology

Cited by 2 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献