SURFACE DOSE ESTIMATION BY A KAP METER FOR KILOVOLTAGE X-RAY BEAMS

Author:

Araki F1ORCID,Umeno S2

Affiliation:

1. Department of Health Sciences, Faculty of Life Sciences, Kumamoto University, 4-24-1 Kuhonji, Chuo-ku, Kumamoto 862-0976, Japan

2. SAGA Heavy Ion Medical Accelerator in Tosu, 3049 Harakogamachi, Tosu, Saga, 841-0071, Japan

Abstract

Abstract This study aims to estimate the entrance surface dose (ESD) of a water phantom for kilovoltage x-ray beams using an air kerma area product meter (KAP meter) equipped in an x-ray unit. The KAP meter was calibrated in terms of the ESD determined by a plane-parallel ionization chamber based on a 60Co absorbed dose-to-water calibration coefficient, ${N}_{D,w}^{{}^{60}\mathrm{C}\mathrm{o}}$. The ESD measured using the KAP meter was verified by comparing it with that estimated by the air kerma calibration coefficient, NK, for x-ray beam qualities. The ratio of ESDs based on ${N}_{D,w}^{{}^{60}\mathrm{C}\mathrm{o}}$ and NK was 1.003 on average and independent of the beam quality. The ESD by the KAP meter was an agreement within ±1.5% with that measured using the plane-parallel chamber for 10 × 10–30 × 30 cm2 fields with a source-surface distance of 75–150 cm. It was possible to estimate the ESD directly in a water phantom for x-ray beams without correction factors compared to the existing air kerma calibration, using a KAP meter calibrated based on ${N}_{D,w}^{{}^{60}\mathrm{C}\mathrm{o}}$.

Funder

Ministry of Education, Culture, Sports, Science, and Technology

Publisher

Oxford University Press (OUP)

Subject

Public Health, Environmental and Occupational Health,Radiology Nuclear Medicine and imaging,General Medicine,Radiation,Radiological and Ultrasound Technology

Reference19 articles.

1. Radiological Protection in Cardiology;International Commission on Radiological Protection,2013

2. Avoidance of Radiation Injuries from Medical Interventional Procedures;International Commission on Radiological Protection,2000

3. Skin injuries from fluoroscopically guided procedures: Part 1, Characteristics of radiation injury;Koenig;AJR.,2001

4. Skin injuries from fluoroscopically guided procedures: Part 2, Review of 73 cases and recommendations for minimizing dose delivered to patient;Koenig;AJR.,2001