Abstract
Abstract
In this study, the development of a reasonable index of optimized exposure dose is attempted. Using a direct-type flat-panel system, noise factors contained in the image are analyzed based on the relative standard deviation method, and it is verified that the proposed index conforms with the appropriate standards regarding minimum exposure dose. The findings indicate that Poisson noise is dominant in the general clinical dose range; this noise fraction formed 90% of the total noise in the system considered in this study. Considering the results of previous surveys, it is estimated that the point at which the content rate of the Poisson noise starts to decrease can be considered as the lower limit, below which the dose cannot be reduced further without compromising on image quality. As the ‘index’ obtained by this method can be easily measured, it is useful as a quantitative indicator for dose determination and optimization.
Publisher
Oxford University Press (OUP)
Subject
Public Health, Environmental and Occupational Health,Radiology, Nuclear Medicine and imaging,General Medicine,Radiation,Radiological and Ultrasound Technology
Reference34 articles.
1. Diagnostic performance of digital versus film mammography for breast-cancer screening;Pisano;N. Engl. J. Med.,2005
2. Cost-effectiveness of digital mammography breast cancer screening;Ann. Intern. Med.,2008
3. Breast cancer screening with imaging: recommendations from the Society of Breast Imaging and the ACR on the use of mammography, breast MRI, breast ultrasound, and other technologies for the detection of clinically occult breast cancer;Lee;J. Am. Coll. Radiol.,2010
4. The 2007 recommendations of the international commission on radiological protection;The International Commission on Radiological Protection;ICRP Publication,2007