Simultaneous transcatheter treatment of ascending aortic aneurysm with aortic and mitral regurgitation: an in vitro study

Author:

Zhao Yuxi1,Li Tao12ORCID,Wu Mingwei1,Zeng Zhaoxiang1,Gao Minxin1,Bao Xianhao1,Xu Ziyi1,Zhou Jian1,Feng Jiaxuan1,Feng Rui1,Jing Zaiping1

Affiliation:

1. Department of Vascular Surgery, Changhai Hospital, Endovascular Diagnosis and Treatment Center for Heart Valvular Diseases, and Endovascular Diagnosis and Treatment Center for Aortic Dissection, Navy Medical University, Shanghai, China

2. Department of Cardiovascular Surgery, Jinling Hospital, Medical School of Nanjing University, Nanjing, Jiangsu, China

Abstract

Abstract OBJECTIVES We sought to explore the efficacy of the endovascular repair of an ascending aortic aneurysm with aortic and mitral regurgitation by 2 novel valved stents. METHODS We established models of ascending aortic aneurysms combined with aortic and mitral regurgitation in 10 pig hearts, then implanted self-expanding aortic fenestrated and mitral valved stents via the transapical approach. We applied a fluoroscopy-guided in vitro setting to test the approach, then analysed continuous circulating flushing at 37°C. Finally, we determined operating times, echocardiography and changes of coronary flow as well as fenestration alignment with the coronary ostia. RESULTS This approach resulted in a 100% overall technical success rate, excellent handling properties and precise positioning. The time taken to implant the 2 valved stents was 59 ± 12 min. Flow of the left and right coronary arteries did not significantly decrease after the stents were implanted (330.4 ± 12.06 ml/min vs 289.4 ± 5.29 ml/min, P < 0.001; 376.8 ± 10.5 ml/min vs 350.0 ± 14.5 ml/min; P < 0.001). We found no obvious regurgitation and perivalvular leakage; nor did the gradients of the aortic and mitral valves as well as of the left ventricular outflow tract increase significantly. The final angiographic examination and profile of the coronary opening confirmed the good position of the valved stents, the exclusion of the aneurysm and the patency of both coronary arteries. CONCLUSIONS These findings indicate the potential for combined transcatheter aortic root and mitral valve replacement in treating aortic root pathologies. In future, in vivo studies are expected to validate this approach and ascertain its durability.

Funder

National Natural Science Foundation of China

Science and Technology Innovation Plan of Shanghai Science and Technology Commission

Shanghai Rising-Star Program

Publisher

Oxford University Press (OUP)

Subject

Cardiology and Cardiovascular Medicine,Pulmonary and Respiratory Medicine,Surgery

Cited by 1 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

1. Assessment of Large Animal Vascular Dimensions for Intra-Aortic Device Research and Development: A Systematic Review;Innovations: Technology and Techniques in Cardiothoracic and Vascular Surgery;2023-03