Late inflow or outflow obstruction requiring surgical intervention after HeartMate 3 left ventricular assist device insertion

Author:

Kaku Yuji1ORCID,Naka Yoshifumi1,Witer Lucas1,Topkara Veli2,Colombo Paolo2,Sayer Gabriel2,Uriel Nir2,Takeda Koji1

Affiliation:

1. Division of Cardiac Surgery, Department of Surgery, Columbia University Irving Medical Center, New York, NY, USA

2. Division of Cardiology, Department of Medicine, Columbia University Irving Medical Center, New York, NY, USA

Abstract

Abstract Inflow or outflow obstruction is a rare, but critical complication after HeartMate 3 left ventricular assist device implantation. We retrospectively reviewed 166 patients who underwent primary HeartMate 3 implantation at our centre between November 2014 and July 2019. Three cases of inflow obstruction and 3 cases of outflow obstruction were identified. Presence of low flow alarm was the most common presenting sign and they were free from heart failure symptoms except in 1 case. They all underwent surgical correction and the median time from primary implantation to reintervention was 406.5 days. Computed tomography angiogram identified a problem in 5 out of 6 cases. Clinical symptoms and/or alarms resolved after surgery, but 1 case developed recurrent low flow alarm due to partial occlusion of inflow by fibrous tissue who eventually underwent heart transplant. All patients remain alive with median follow-up of 433.5 days. Prompt surgical intervention in late inflow and outflow obstruction after HeartMate 3 implantation resulted in favourable outcomes.

Publisher

Oxford University Press (OUP)

Subject

Cardiology and Cardiovascular Medicine,Pulmonary and Respiratory Medicine,Surgery

Reference10 articles.

1. Two-year outcome with a magnetically levitated cardiac pump in heart failure;Mehra;N Engl J Med,2018

2. A tale of the twist in the outflow graft: an analysis from the MOMENTUM 3 trial;Mehra;J Heart Lung Transplant,2018

3. Outflow graft Occlusion of the HeartMate 3 left ventricular assist device;Duero Posada;Circ Heat Fail,2017

4. A case of an obstructive inflow thrombus in a HeartMate 3 from the left ventricle into the pump;Eulert-Grehn;J Heart Lung Transplant,2018

5. A single center experience: four cases of late twisting of outflow grafts in HeartMate 3;Mueller;J Heart Lung Transplant,2018

Cited by 3 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献