An in vitro comparison of internally versus externally mounted leaflets in surgical aortic bioprostheses

Author:

Vriesendorp Michiel D1ORCID,de Lind van Wijngaarden Rob A F1,Rao Vivek2ORCID,Moront Michael G3,Patel Himanshu J4,Sarnowski Edward5,Vatanpour Sepehr5,Klautz Robert J M1ORCID

Affiliation:

1. Department of Cardiothoracic Surgery, Leiden University Medical Center, Leiden, Netherlands

2. Department of Cardiovascular Surgery, Toronto General Hospital, Toronto, ON, Canada

3. Department of Cardiothoracic Surgery, ProMedica Toledo Hospital, Toledo, OH, USA

4. Department of Cardiac Surgery, University of Michigan, Ann Arbor, MI, USA

5. Research & Development, Medtronic, Santa Ana, CA, USA

Abstract

Abstract OBJECTIVES To improve haemodynamic performance, design modifications of prosthetic valves have been proposed with each new generation of valves. These different designs also impact the amount of mechanical wear, because mechanical stresses are distributed differently. Because long-term evidence for new prosthetic valves is lacking, this in vitro study compared hydrodynamic performance and durability among 3 currently available bioprosthetic valves with internally (IMLV) or externally mounted leaflets (EMLV). METHODS Prostheses of the internally mounted Medtronic Avalus and Carpentier-Edwards Perimount Magna Ease valves were compared to prostheses of the externally mounted Abbott Trifecta valve. For each labelled size (e.g. 19, 21 and 23) of the 3 types, 3 valves underwent accelerated wear testing for up to 600 million cycles, corresponding to ∼15 years of simulated wear. The valves underwent hydrodynamic testing and visual inspection. RESULTS EMLV had the largest effective orifice area and lowest pressure gradient for each labelled size at baseline and 600 million cycles; the effective orifice area and the pressure gradient were equivalent for the 2 types of IMLV. Five of 9 EMLVs had at least 1 hole or tear in the leaflet tissue around the stent posts, which resulted in severe regurgitation at 500 million cycles in 2 cases. All IMLVs were intact at 600 million cycles with minimal tissue wear. CONCLUSIONS EMLV showed superior hydrodynamic performance but inferior mechanical durability compared to IMLV after 600 million cycles of testing. The primary failures were because of significant mechanical abrasion in the commissural region, which may warrant close monitoring of EMLV during long-term follow-up.

Publisher

Oxford University Press (OUP)

Subject

Cardiology and Cardiovascular Medicine,Pulmonary and Respiratory Medicine,Surgery

Cited by 24 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3