Affiliation:
1. Cardiothoracic Department, Maastricht University Hospital, Maastricht, Netherlands
2. Cardiothoracic Department, Brussels University Hospital, Brussels, Belgium
Abstract
Abstract
OBJECTIVES
The aim of this study was to evaluate the lesion size and depth of radiofrequency (RF) ablation in a simultaneous biparietal bidirectional bipolar (SBB) approach, compared to a simultaneous and staged unipolar and uniparietal bipolar setup [simultaneous uniparietal bipolar (SiUB) and staged uniparietal bipolar (StUB), respectively].
METHODS
Fresh left atrial porcine tissue was mounted into the ABLA-BOX simulator. Different ablation approaches were tested: (i) SBB: a concept consisting of SBB endo-epicardial ablation, (ii) SiUB: simultaneous epicardial uniparietal bipolar and endocardial unipolar ablation and (iii) StUB: staged epicardial uniparietal bipolar and endocardial unipolar ablation. In the StUB setup, a 1-h interval between the epi-endo ablation was respected.
RESULTS
Transmural lesions were present in 90% of the bipolar biparietal ablations, yet no full transmurality was observed in the simultaneous nor in the staged unipolar with uniparietal bipolar ablation group. In SBB, the area and volume of the ablation lesions were smaller (523.33 mm2/mm and 52.33 mm3/mm, respectively) than in SiUB (588.17 mm2/mm and 58.81 mm3/mm, respectively) and StUB (583.76 mm2/mm and 58.37 mm3/mm, P = 0.044). Also, in SBB, the overall, epicardial and endocardial maximum diameters of the lesions (1.59, 1.57 and 1.52 mm; respectively) were smaller than in SiUB (2.38, 2.26 and 2.33 mm; respectively) and in StUB (2.36, 2.28 and 2.14 mm; respectively, all P < 0.001).
CONCLUSIONS
Although bipolar biparietal bidirectional RF ablation results in smaller lesions than uniparietal bipolar and unipolar ablation, their capacity to penetrate the tissue is much higher. Moreover, in uniparietal RF applications, the energy spreads in the superficial layers of the tissue but fails to penetrate. Therefore, the degree of transmurality is much higher when using such a ‘truly bipolar’ ablation approach.
Publisher
Oxford University Press (OUP)
Subject
Cardiology and Cardiovascular Medicine,Pulmonary and Respiratory Medicine,Surgery
Cited by
1 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献