A randomized cross-over study with objective quantification of the performance of an asymmetric triacetate and a polysulfone dialysis membrane using different anticoagulation strategies

Author:

Vanommeslaeghe Floris1,Josipovic Iván2,Boone Matthieu2,Dhondt Annemie1,Van Biesen Wim1ORCID,Eloot Sunny1

Affiliation:

1. Nephrology Department, Ghent University Hospital, Ghent, Belgium

2. Centre for X-ray Tomography, Physics and Astronomy, Ghent University, Ghent, Belgium

Abstract

Abstract Background Different strategies can be used to counteract coagulation of extracorporeal systems. Systemic anticoagulation is most widely used in routine clinical practice, but can be contraindicated in specific settings. The Solacea™ dialyser, containing the asymmetric triacetate membrane, claims improved biocompatibility, which should result in decreased tendency for coagulation. We quantified the performance of the Solacea™ versus the FX800CORDIAX dialyser regarding resistance to fibre blocking as assessed by micro-computed tomography (CT). Methods This cross-over study with four arms randomized consecutively 10 maintenance haemodialysis patients to a 4-h post-dilution haemodiafiltration session at midweek, using either Solacea™ 19 H or FX800CORDIAX, with either regular or half dose of anticoagulation (EC2017/1459-NCT03820401). Dialyser fibre blocking was visualized in the dialyser outlet potting using a 3D CT scanning technique on micrometre resolution. Extraction ratios of middle molecules [myoglobin, lambda and kappa free light chains (FLCs)] were determined. Results The relative number of open fibres post-dialysis was lower in FX800CORDIAX versus Solacea™ dialyser, and this was irrespective of the anticoagulation dose used or the threshold for counting open fibres. Extraction ratios of FLCs were not different at regular anticoagulation between Solacea™ and FX800CORDIAX (21% ± 4% for kappa and 32% ± 8% for lambda with Solacea™ versus 23% ± 7% and 38% ± 6% for FX800CORDIAX), but were superior with the Solacea™ (34% ± 12% versus 22% ± 8% with FX800CORDIAX; P = 0.02) for myoglobin in case of halving anticoagulation dose. No clinically relevant albumin loss was detected. Conclusions The Solacea™ dialyser seems to be promising for use in conditions where systemic anticoagulation is contraindicated, as even under conditions of low systemic anticoagulation, virtually no signs of fibre blocking could be observed using the sensitive micro-CT scanning technique. This finding is in line with its presumed good performance in terms of biocompatibility.

Funder

Ghent University Hospital KOF 2015 grant for F.V., Ghent University

unrestricted grant from Nipro Corporation

Publisher

Oxford University Press (OUP)

Subject

Transplantation,Nephrology

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3