Comparative Evaluation of Carboxyhemoglobin Quantification in Postmortem Whole Blood by CO-Oximetry and Headspace Gas Chromatography with Flame Ionization Detection and Atom Absorption Spectrophotometry

Author:

Kristoffersen Lena1ORCID,Vevelstad Merete1,Hansen Inger-Ann1,Strømsvåg Berit1,Strand Dag Helge1

Affiliation:

1. Department of Forensic Sciences, Oslo University Hospital , Lovisenberggt. 6, Oslo 0456, Norway

Abstract

Abstract A comparative evaluation of two methods used for carboxyhemoglobin (COHb) determination in postmortem whole blood was performed: carbon monoxide (CO)-oximetry measuring at 128 wavelengths and headspace gas chromatography with flame ionization detection (HS-GC--FID) where CO was determined after catalytic reduction of CO to CH4 and Fe was determined by atom absorption spectrophotometry (AAS, 248.3 nm). An aliquot of 100 µL whole blood was loaded into the CO-oximetry module. In the HS-GC--FID analysis, to 1.0 mL of whole blood, 3.0 mL of saponin solution was added, mixed and centrifuged. To 20 mL HS vials, 400 µL of the supernatant was added and the vials were immediately sealed. One milliliter of potassium hexacyanoferrat (III) solution was added through the HS septum and mixed. The samples were incubated at 70°C for 5 min. CO was separated using He as carrier gas and a CP-Molsieve 5 Å PLOT capillary column. Fe was determined using 400 µL of the saponin supernatant diluted to 10 mL by water. During a period of ∼3 years, 124 postmortem whole blood samples were analyzed. Bland–Altman method comparison showed satisfactory agreement and no significant bias between the methods for the whole saturation range (5 to 85% COHb). Five samples, all with %COHb >40, showed deviations of more than 10% COHb in absolute terms. One sample, in the lower COHb range <10%, was false negative on the CO-oximetry method. The between-assay accuracy, reported as bias, at 60% COHb was −0.8% and −9.0%, and precision, reported as relative standard deviation, was 1.6% and 7.7%, for the CO-oximetry and HS-GC--FID-AAS methods, respectively. Both methods obtained satisfactory results in proficiency testing rounds, with z-scores <±2 (n = 11). This study showed that the CO-oximetry method based on the 128-wavelength principle and the HS-GC--FID-AAS method are comparable and satisfactory for %COHb determination in postmortem whole blood.

Publisher

Oxford University Press (OUP)

Subject

Chemical Health and Safety,Health, Toxicology and Mutagenesis,Toxicology,Environmental Chemistry,Analytical Chemistry

Reference22 articles.

1. Mortality associated with exposure to carbon monoxide in WHO European member states;Braubach;Indoor Air,2013

2. Suicide from carbon monoxide poisoning in South Korea: 2006-2012;Choi;Journal of Affective Disorders,2014

3. Carbon monoxide poisoning deaths in the United States, 1999 to 2012;Sircar;American Journal of Emergency Medicine,2015

4. National unintentional carbon monoxide poisoning estimates using hospitalization and emergency department data;Stearns;American Journal of Emergency Medicine,2019

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3