A case report of entrapment of PentaRay catheter in the mechanical mitral valve: a known complication experienced anew

Author:

Wei Hui-Qiang1,Liu Fangzhou1,Xue Yumei1,Wu Shulin1,Zhan Xianzhang1ORCID

Affiliation:

1. Department of Cardiology, Guangdong Cardiovascular Institute, Guangdong Provincial People's Hospital, Guangdong Academy of Medical Sciences , No. 106 Zhongshan 2nd Road, Yuexiu District, 510000 Guangzhou , China

Abstract

Abstract Background Complex atrial tachycardia (AT) is commonly observed in patients with cardiac surgery. High-density mapping is widely adopted for catheter ablation of complex AT in patients with cardiac surgery. Several case reports have described that PentaRay mapping catheter can be trapped in the mechanical valve and sheared off and successful retrieval of the spline by a snare system. We described a rare case in which PentaRay mapping catheter spline was successfully retrieved from the distal anterior tibial artery by direct syringe suction via the diagnostic catheter following entrapment in the mechanical mitral valve (MV) and rupture of the spline. Case summary A 70-year-old female with mechanical bileaflet MV underwent catheter ablation for AT. During mapping in left atrium, the catheter was entrapped in mechanical MV and sheared off. We attempted to release the entrapped the spline by advancing the ablation catheter towards the stuck disc and pushing on the hinge portion of the disc with the catheter tip. The stuck and closed disc was opened, and the deeply entrapped spline was released. However, the entrapped PentaRay spline floated through the Valsalva sinus and strayed into the distal left anterior tibial artery. Fortunately, we successfully retrieved the spline from the distal anterior tibial artery by direct syringe suction instead of a snare system. Discussion The possibility of the entrapment and subsequent rupture of the spline should always be considered during mapping the site close to mechanical valve. A rapid retrieval of embolized material should be carried out. If the spline strays into the distal and small artery in which the snare system is difficult to advance, a direct syringe suction via the diagnostic catheter may be attempted.

Publisher

Oxford University Press (OUP)

Reference4 articles.

1. Lesional tachycardias related to mitral valve surgery;Markowitz;J Am Coll Cardiol,2002

2. A novel technique to release a PentaRay entrapped in a mechanical mitral valve using an ablation catheter;Yagishita;JACC Clin Electrophysiol,2020

3. Multipolar catheter entrapment in a mechanical mitral valve and successful retrieval of a sheared spine straying into the coronary artery;Asahina;HeartRhythm Case Rep,2023

4. How to release PentaRay catheter entrapped in the hinge point of mechanical mitral valve?;Kawaji;Europace,2020

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3