Riata lead failure manifests as electrical storm with lead parameters showing unique postural variations: a case report

Author:

Uppal Abhimanyu1ORCID,Kathuria Sanjeev1,Shah Bhushan1ORCID,Trehan Vijay1

Affiliation:

1. Department of Cardiology, GB Pant Institute of Post Graduate Education and Research, MAMC, JLN Road, New Delhi 110002, India

Abstract

Abstract Background Riata implantable cardioverter-defibrillator (ICD) leads are prone to a unique type of mechanical lead failure causing conductor externalization (CE) which may be complicated by a delayed-onset electrical lead failure (ELF). Case summary A 60-year-old male with symptomatic, severe ischaemic cardiomyopathy, and atrial fibrillation following a prior anterior wall myocardial infarction received a dual-chamber ICD with 7F-RiataST ventricular lead as a primary prevention strategy against sudden cardiac death in 2008. In 2017, a pulse generator replacement was performed for elective replacement indicator status. At that time, CE was noted in the ventricular lead but the electrical lead parameters were normal, hence lead replacement was decided against and the patient was closely followed up thereafter. Four years later, the patient presented with multiple ICD shocks within 48 h. Implantable cardioverter-defibrillator interrogation showed noise on the ventricular electrogram (EGM) channel that was detected as ventricular fibrillation (VF) episodes, triggering inappropriate ICD therapy (five ICD detected VF events within 24 h triggering three antitachycardia pacing therapies and one shock). Lead impedance and R-wave amplitude were within normal range in supine position but dramatically worsened in sitting posture. A new ventricular lead was implanted and the old lead abandoned. The patient has not experienced any device therapy in the follow-up period. Discussion An electrically inert CE of Riata ICD leads needs close follow-up because an ELF may occur even after several years. A careful analysis of EGMs including postural changes in lead parameters can aid in detection and better characterization of underlying electrical dysfunction following CE.

Publisher

Oxford University Press (OUP)

Subject

Cardiology and Cardiovascular Medicine

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3