Affiliation:
1. Department of Medicine, Duke University Hospital, Durham, NC, USA
Abstract
Abstract
Background
With the ongoing coronavirus disease 2019 (COVID-19) epidemic, remote monitoring of patients with implanted cardiac devices has become more important than ever, as physical distancing measures have placed limits on in-clinic device monitoring. Remote monitoring alerts, particularly those associated with heart failure trends, have proved useful in guiding care in regard to monitoring fluid status and adjusting heart failure medications.
Case summary
This report describes use of Boston Scientific’s HeartLogic algorithm, which is a multisensor device algorithm in implantable cardioverter-defibrillator devices that is proven to be an early predictor of heart failure decompensation by measuring several variables, including respiratory rate, nighttime heart rate, and heart sounds. We present three cases of patients who were actively surveilled by the various HeartLogic device algorithm sensors and were identified to have increasing respiratory rates high enough to trigger a HeartLogic alert prior to a positive COVID-19 diagnosis.
Discussion
We propose that the HeartLogic algorithm and its accompanying individual physiologic sensors demonstrate potential for use in identifying non-heart failure-related decompensation, such as COVID-19-positive diagnoses.
Publisher
Oxford University Press (OUP)
Subject
Cardiology and Cardiovascular Medicine
Cited by
4 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献