A case report of an unusual cause of increased impedance after generator exchange

Author:

Feldman Daniel R12ORCID,Kiessling Mark H2,Sangha Rajbir S12ORCID

Affiliation:

1. Department of Medicine, Geisel School of Medicine, Dartmouth College , 1 Rope Ferry Road, Hanover, New Hampshire 03755 , USA

2. Heart & Vascular Center, Dartmouth-Hitchcock Medical Center , 1 Medical Center Drive, Lebanon, New Hampshire 03756 , USA

Abstract

Abstract Background Remote monitoring has emerged as a complement to in-person care for patient with cardiac implantable electronic devices (CIEDs). It provides the care team with information about device integrity, programming issues, or other medical data (i.e. arrhythmias) and since 2015 has been recognized as a part of standard management by the Heart and Rhythm Society for all patients with CIEDs. However, while it can provide invaluable information to providers, the volume of generated data can increase the risk of oversight. We present a novel case of apparent device malfunction that on closer scrutiny was obvious, but provides a lesson in the mechanisms by which data can be artifactual. Case summary A 62-year-old male presented after his cardiac resynchronization therapy-defibrillator (CRT-D) alerted him that his device was at an elective replacement interval (ERI). He underwent an uncomplicated generator exchange; however, 2 weeks later, a remote alert showed that his device was at ERI and all impedances were above the upper limit. Device interrogation the following day demonstrated that the new device was functioning appropriately and his home monitor had in fact paired with his old generator. He obtained a new home monitor, and subsequent remote transmissions have demonstrated that his device is functioning appropriately. Discussion This case demonstrates the importance of careful review of details from home-monitoring data. While concerning for device malfunction, there could be alternative causes when alerts are generated by remote monitoring. To our knowledge, this is the first report of this mechanism of alert via a home-monitoring device and should be considered when reviewing unusual remote download data.

Publisher

Oxford University Press (OUP)

Reference7 articles.

1. Impedance in the diagnosis of lead malfunction;Swerdlow;Circ Arrhythm Electrophysiol,2020

2. A case series of anomalous high pacing lead impedances in normally functioning leads;Mourad;Heart Rhythm Case Rep,2015

3. Abrupt increase in impedance measurements as detected via remote monitoring: what is the cause?;Nayeri;Heart Rhythm Case Rep,2015

4. Rise in ICD shock impedance: lead fracture or death?;Stroobandt;Pacing Clin Electrophysiol,2012

5. Long-term patient acceptance of and satisfaction with implanted device remote monitoring;Ricci;Europace,2010