Abstract
Abstract
Confocal micro-beam X-ray fluorescence analysis (confocal micro-XRF) is a non-destructive analytical tool for investigation of sample composition that enables acquiring three-dimensionally resolved information. This work describes a calibration procedure of a laboratory confocal micro-XRF setup, which leads to determination of its characteristic parameters. The calibration is performed using a tabletop confocal micro-XRF spectrometer designed recently at the Czech Technical University in Prague. The calibration procedure performed within this work comprises the essential steps of the setup characterization: excitation spectrum calculation, experimental determination of energy-dependent confocal volume size and integral sensitivity and calculation of the spectrometer sensitivity function. The results of the setup calibration will be used for development of a procedure enabling quantitative evaluation of the measured depth profiles.
Publisher
Oxford University Press (OUP)
Subject
Public Health, Environmental and Occupational Health,Radiology Nuclear Medicine and imaging,General Medicine,Radiation,Radiological and Ultrasound Technology