ESTABLISHMENT OF LOCAL DIAGNOSTIC REFERENCE LEVELS FOR COMMON PROCEDURES OF COMPUTED TOMOGRAPHY IN YAZD PROVINCE

Author:

Zamani Hamed1,Masjedi Hamidreza1,Omidi Reza1,Hosein Zare Mohammad1

Affiliation:

1. Medical Physics Department, School of Medicine, Shahid Sadoughi University of Medical Sciences and Health Services, Yazd, Yazd Province, Iran

Abstract

Abstract Objective: The aim of this study was to propose first established diagnostic reference levels (DRLs) in computed tomography (CT) for adults, based on volume-averaged CTDI and dose length product (DLP) metrics in Yazd Province. Materials and Methods: Six multislice CT scanners located at diverse areas of Yazd Province and seven common procedures were selected for the present study. For each procedure, at least twenty patients 18 years and older were sampled at each institution. For each patient, dose report data and scan parameters as well as patient’s information were abstracted from picture archiving and communication system. Results: Proposed DRLs in terms of computed tomography dose index (mGy) and DLP (mGy.cm) were as follows: brain (42, 527), sinus (25, 220), neck (14, 264), abdomen-pelvis (11, 295), routine chest (8, 247), CT pulmonary angiogram (32, 261) and chest HRCT (11, 455), respectively, slightly lower compared to other investigations. Conclusion: The proposed DRLs in this study should be considered as the local DRLs for the seven most common adult CT examinations in Yazd province so as to optimize the patient dose while maintaining acceptable image quality for the clinical task.

Funder

Shahid Sadoughi University of Medical Sciences and Health Services

Publisher

Oxford University Press (OUP)

Subject

Public Health, Environmental and Occupational Health,Radiology, Nuclear Medicine and imaging,General Medicine,Radiation,Radiological and Ultrasound Technology

Reference28 articles.

1. National survey of doses from CT in the UK: 2003;Shrimpton;Br. J. Radiol.,2006

2. National Council on Radiation Protection and Measurements report shows substantial medical exposure increase;Schauer;Radiological Society of North America, Inc.,2009

3. The measurements of CTDIw and DLP in routine CT protocols and comparison with reference dose level in Kashan, Iran;Ali Asgharzadeh;Iran. J. Radiat. Saf. Meas.,2014

4. CT scanner dosimetry;Shrimpton;Br. J. Radiol.,1998

5. CT radiation dose optimization and estimation: an update for radiologists;Goo;Korean J. Radiol.,2012