Affiliation:
1. U.S. Food and Drug Administration, Division of Drug Analysis, St. Louis, MO 63101
Abstract
Abstract
A simple, isocratic liquid chromatographic method for assay of thalidomide in tablets, capsules, and raw materials was developed. The method uses a Nova-Pak octadecylsilane bonded-phase column (150 × 3.9 mm, 4 μm particle size), a mobile phase of acetonitrile-water (15 + 85), a flow rate of 1 mL/min, detection at 237 nm, and phenacetin as internal standard. Phosphoric acid was used in preparation of sample solutions to inhibit thalidomide hydrolysis. Assays ranged from 99.3 to 100.4% in raw materials from 4 manufacturers, from 79.7 to 104.8% in tablets from 7 manufacturers, and from 75.3 to 102.6% in capsules from 4 manufacturers. Assay method precisions for triplicate analyses on 5 days were 0.30% for tablets, 0.22% for capsules, and 0.22% for raw materials. Recovery from simulated tablet formulations was 100%. The method has been used to analyze individual tablets and capsules for determination of content uniformity.
Publisher
Oxford University Press (OUP)
Subject
Pharmacology,Agronomy and Crop Science,Environmental Chemistry,Food Science,Analytical Chemistry
Cited by
4 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献