Affiliation:
1. Asahi University Hospital, Department of Radiation Oncology, Gifu, Japan
2. Gifu University Hospital, Department of Radiology, Gifu, Japan
Abstract
Abstract
Intensity-modulated radiotherapy (IMRT) has been used for breast cancer as well as in field-in-field techniques. Few dosimetric comparison studies have been conducted using IMRT and volumetric modulated arc therapy (VMAT) for Japanese patients. We aimed to study such patients. Thirty-two patients with left-sided breast cancer were enrolled. We conducted the following five treatment plans: two field-static IMRT (2F-S-IMRT), four field-static IMRT (4F-S-IMRT), 40° dual partial arc VMAT (40d-VMAT), 80° dual partial arc VMAT (80d-VMAT) and 210° partial VMAT (210p-VMAT). We evaluated the following: level of coverage of planning target volume (PTV) of 95% for irradiation at a dose of 50 Gy (D95) and the percentage of the heart and left anterior descending artery (LAD) volume that received 10 Gy or more (V10). As a result, the coverage of 40d-VMAT for the prescribed PTV dose of D95 was significantly lower than that of the other treatment plans (P < 0.05). Regarding heart V10 and LAD V10, 2F-S-IMRT, 40d-VMAT and 80d-VMAT showed significantly lower dose than the other treatment plans (P < 0.05). In conclusion, among the five plans, 2F-S-IMRT is recommended for Japanese patients because of high coverage of D95 of PTV, low V10 of the heart and LAD and the monitor unit value was the lowest.
Publisher
Oxford University Press (OUP)
Subject
Health, Toxicology and Mutagenesis,Radiology, Nuclear Medicine and imaging,Radiation
Cited by
5 articles.