Performance of three iGrav superconducting gravity meters before and after transport to remote monitoring sites

Author:

Schäfer Florian1,Jousset Philippe1,Güntner Andreas12,Erbas Kemal1,Hinderer Jacques3,Rosat Séverine3ORCID,Voigt Christian1,Schöne Tilo1ORCID,Warburton Richard4

Affiliation:

1. Helmholtz Centre Potsdam GFZ German Research Centre for Geosciences, Potsdam, Germany

2. University of Potsdam, Institute of Environmental Science and Geography, Potsdam, Germany

3. IPGS/EOST, CNRS/Université of Strasbourg, France

4. GWR Instruments, Inc., San Diego, CA, USA

Abstract

SUMMARY High spatial and temporal resolution of gravity observations allows quantifying and understanding mass changes in volcanoes, geothermal or other complex geosystems. For this purpose, accurate gravity meters are required. However, transport of the gravity meters to remote study areas may affect the instrument's performance. In this work, we analyse the continuous measurements of three iGrav superconducting gravity meters (iGrav006, iGrav015 and iGrav032), before and after transport between different monitoring sites. For 4 months, we performed comparison measurements in a gravimetric observatory (J9, Strasbourg) where the three iGravs were subjected to the same environmental conditions. Subsequently, we transported them to Þeistareykir, a remote geothermal field in North Iceland. We examine the stability of three instrumental parameters: the calibration factors, noise levels and drift behaviour. For determining the calibration factor of each instrument, we used three methods: First, we performed relative calibration using side-by-side measurements with an observatory gravity meter (iOSG023) at J9. Secondly, we performed absolute calibration by comparing iGrav data and absolute gravity measurements (FG5#206) at J9 and Þeistareykir. Thirdly, we also developed an alternative method, based on intercomparison between pairs of iGravs to check the stability of relative calibration before and after transport to Iceland. The results show that observed changes of the relative calibration factors by transport were less than or equal to 0.01 per cent. Instrumental noise levels were similar before and after transport, whereas periods of high environmental noise at the Icelandic site limited the stability of the absolute calibration measurements, with uncertainties above 0.64 per cent (6 nm s–2 V–1). The initial transient drift of the iGravs was monotonically decreasing and seemed to be unaffected by transport when the 4K operating temperatures were maintained. However, it turned out that this cold transport (at 4 K) or sensor preparation procedures before transport may cause a change in the long-term quasi-linear drift rates (e.g. iGrav015 and iGrav032) and they had to be determined again after transport by absolute gravity measurements.

Funder

French National Research Agency

German Federal Ministry

BMBF

Helmholtz Centre Potsdam - GFZ German Research Centre for Geosciences

Landsvirkjun

Publisher

Oxford University Press (OUP)

Subject

Geochemistry and Petrology,Geophysics

Reference40 articles.

1. Earth tides;Agnew,2015

2. Three year comparison between a superconducting gravimeter (GWR C026) and an absolute gravimeter (FG5# 206) in Strasbourg (France);Amalvict;J. Geod. Soc. Jpn.,2001

3. Analysis of co-located measurements made with a LaCoste&Romberg Graviton-EG gravimeter and two superconducting gravimeters at Strasbourg (France) and Yebes (Spain);Arnoso;Acta Geod. Geophys.,2014

4. Time stability of spring and superconducting gravimeters through the analysis of very long gravity records;Calvo;J. Geodyn.,2014

5. Tidal spectroscopy from a long record of superconducting gravimeters in Strasbourg (France);Calvo,2016

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3