Validation of krypton as a new tracer gas for the standardization tests of collective and individual protection systems

Author:

Cabaset Elise1,Marchal Mathieu1,Santandrea Audrey1,Herbinet Olivier2,Belut Emmanuel1,Lechêne Sullivan1,Marsteau Stéphanie1,Leclerc Jean-Pierre1

Affiliation:

1. Département d’Ingénierie des Procédés, INRS , 1 rue du Morvan, CS 60027 54519 Vandoeuvre les Nancy Cedex , France

2. LRGP, CNRS, Université de Lorraine , F-54000 Nancy , France

Abstract

Abstract Sulfur hexafluoride (SF6) is the reference tracer gas in many international standards for characterizing respiratory protective devices (RPD), fume cupboards, building ventilations, and other installations. However, due to its significant impact on global warming, its use is becoming increasingly restrictive. Krypton 84 (Kr) was chosen to be a possible replacement based on theoretical and practical criteria for the properties that a substitute gas should possess. While compliance with these criteria is generally sufficient to guarantee the reliability of the choice, it is essential in the case of widespread use such as a standard to validate experimentally that this tracer has the same behavior as SF6. In this regard, numerous tests have been carried out to characterize the face leakage of RPD and the rupture of containment of fume cupboards performance tests under different operating conditions. The results obtained are identical with both tracers and lead us to propose the use of Kr as a new reference gas in standards for which SF6 was used.

Publisher

Oxford University Press (OUP)

Subject

Public Health, Environmental and Occupational Health

Reference40 articles.

1. Methods of testing performance of laboratory fume hoods;ANSI/ASHRAE Standard 110-2016,2016

2. Standard test method for determining air change in a single zone by means of a tracer gas dilution;ASTM E0741-11 R17,2017

3. Flow and contaminant transfer at the interface of an aerodynamic containment barrier subjected to the wake of a moving obstacle;Belut,2023

4. Impact of multiple consecutive Donnings on filtering facepiece respirator fit;Bergman,2012

5. Rapid robot-aided on-site testing of fume cupboards;Bittner,2022