Aerosol and surface contamination assessment of a novel ventilated infectious aerosol containment device

Author:

Fang Runcheng1ORCID,Andrus Niles1,Dominguez Thomas1,Sleeth Darrah K1,Jones Rachael M12ORCID

Affiliation:

1. Division of Occupational & Environmental Health, Department of Family & Preventive Medicine, University of Utah , Salt Lake City, UT 84108 , USA

2. Department of Environmental Health Sciences, Fielding School of Public Health, University of California Los Angeles , 650 Charles E Young Dr. S, Los Angeles, CA 90095 , USA

Abstract

Abstract Healthcare personnels (HCPs) are at risk of respiratory infectious diseases during patient care activities. HCPs rely primarily on personal protective equipment to prevent pathogen exposures, but there is a need to develop alternative, or complementary control strategies, including engineering controls. The objective of this study was to evaluate the ability of the 3 designs (denoted D1A, D1B, and D2) of the University of Utah Containment Ventilation for Exposure Reduction (U-COVER), a protective barrier enclosure device to contain respirable aerosols when placed over a simulated patient. The 2 primary performance metrics were the percent reduction in: (i) the concentration of respirable aerosols in the simulated breathing zone of an HCP, and (ii) surface contamination outside the device, which were tested using salt aerosols and fluorescein aerosols, respectively. Briefly, salt or fluorescein aerosols were generated as though expelled by a prone patient under 3 conditions: (i) no device (control), (ii) with the device but without exhaust ventilation, and (iii) with the device with exhaust ventilation. Device D2 was also tested under simulated use conditions, in which cardboard “arms” were placed inside the device ports. All 3 device designs showed the ability to reduce particle concentrations in the simulated HCP breathing zone and on surfaces by >99% with exhaust ventilation compared to the control condition. Without exhaust ventilation, device performance was lower and highly variable. Under simulated use conditions, device D2 reduced particle concentrations in the simulated HCP breathing zone by ≥91% and on surfaces by >99% relative to control for all combinations of “arms” tested. The U-COVER device demonstrates excellent aerosol containment and warrants further testing with dynamic simulated or actual use conditions.

Funder

Department of Defense

Rocky Mountain Center for Occupational and Environmental Health

Publisher

Oxford University Press (OUP)

Subject

Public Health, Environmental and Occupational Health

Reference26 articles.

1. Development of a new Aerosol Barrier Mask for mitigation of spread of SARS-CoV-2 and other infectious pathogens;Abi Karam;Respir Med,2021

2. An aerosol containment and filtration test for intubation during the COVID-19 pandemic;Brant-Zawadski;Surg Innov,2021

3. An operative barrier system for skull base and mastoid surgery: creating a safe operative theatre in the era of covid-19;Cottrell,2020

4. Effective design of barrier enclosure to contain aerosol emissions from COVID-19 patients;Daniel;Indoor Air,2021

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3